Interactive guide · IVDR · 13 min read
IVDR Classification Rules: Find Your IVD's Class (A, B, C or D) with the Annex VIII Picker
The EU In Vitro Diagnostic Medical Devices Regulation (IVDR) sorts every IVD into Class A, B, C or D using seven rules in Annex VIII. The class decides whether a notified body is involved, how deeply it looks and how often you report after launch. Most IVDs now need a notified body, which most didn't under the old directive. This guide explains each rule with examples from the official guidance, and the picker walks you through them for your own product.
Class A
Lowest risk. Self-declared, unless placed on the market sterile.
Class B
Moderate individual risk. The default class when no other rule applies.
Class C
High individual or moderate public health risk, including cancer, genetic and companion diagnostic tests.
Class D
High individual and public health risk, such as blood screening and life-threatening transmissible agents.
How IVDR classification works
Article 47 of the IVDR divides devices into Classes A, B, C and D, taking into account their intended purpose and inherent risks, and says classification follows Annex VIII. Annex VIII has implementing rules, which tell you how to apply the rules, and seven classification rules.[1]
This is a big change from the old IVD Directive, which used fixed lists of products that needed a notified body. Under the IVDR, classification is risk-based, and the Commission estimates that about 80% of IVDs need a notified body, against about 8% under the directive.[1][2]
Intended purpose decides
Rules are applied according to the intended purpose you state. If you state several intended purposes that fall into different classes, the device goes into the higher class.[1]
The highest class wins
You must consider all the rules. If several rules apply to the same device, the one giving the higher class applies.[1]
Each device on its own
A device used with another device is classified separately, and accessories are classified in their own right.[1]
First-line, confirmatory or supplemental
Each rule applies equally to first-line, confirmatory and supplemental assays.[1]
The four classes and what each one means for you
| Class | Examples from MDCG 2020-16 | Conformity assessment | Post-market reporting |
|---|---|---|---|
| A | Clinical chemistry analysers, PCR thermocyclers, specimen receptacles, nucleic acid extraction kits | You draw up the technical documentation and declaration of conformity yourself. If placed on the market sterile, a notified body assesses the sterility aspects. | Post-market surveillance report, updated when necessary |
| B | Electrolyte tests, H. pylori tests, urine dipsticks, pregnancy self-tests, controls without assigned values | Notified body assesses your quality management system and the technical documentation of representative devices. Self-tests and near-patient tests also have their technical documentation assessed. | Post-market surveillance report, updated when necessary |
| C | Cancer screening tests, human genetic tests, companion diagnostics, glucose meters for self-testing, troponin tests | Notified body assesses the quality system and the technical documentation of at least one representative device per generic device group, and of every self-test, near-patient test and companion diagnostic. Companion diagnostics involve a consultation with a medicines authority. | Periodic safety update report (PSUR), updated at least once a year |
| D | Blood donor screening for HIV and hepatitis, ABO grouping, HIV viral load | Notified body assesses the quality system and the technical documentation of every device. Where an EU reference laboratory is designated, it verifies performance and tests batches; some first certifications go to an expert panel. | PSUR, updated at least once a year |
Conformity assessment routes are in Article 48 and Annexes IX to XI; post-market reports in Articles 80 and 81. Class C and D devices also need a summary of safety and performance (Article 29).[1]
The classification picker
Answer the questions for your product. The picker applies the rules that fit and gives you the highest class with the rules behind it. Confirm the result against Annex VIII and MDCG 2020-16, and record your rationale in your technical documentation.
1. What is it?
Some products aren't IVDs at all: products for general laboratory use or research use only, invasive sampling products, internationally certified reference materials and materials for external quality assessment schemes fall outside the IVDR.
2. Does it measure directly through its own sensors, without separate reagents, or is it a meter for self-testing?
2. Tick any that apply (Rule 1 and Rule 2)
3. Tick any that apply (Rule 3)
4. Who performs the test?
If it's intended for both professionals and lay users, choose self-testing.
5. Is it a self-test for pregnancy, fertility or cholesterol, or for glucose, erythrocytes, leucocytes or bacteria in urine?
All seven rules at a glance
Examples are from MDCG 2020-16, the Medical Device Coordination Group's guidance on IVD classification.[1][3]
| Rule | Covers | Class | Examples |
|---|---|---|---|
| 1 | Blood, tissue and organ donor screening for transmissible agents; life-threatening agents with a high risk of spreading; infectious load monitoring for life-threatening diseases | D | Anti-HIV 1/2, HBsAg and anti-HCV donor screening; HIV and HBV viral load |
| 2 | Blood grouping and tissue typing for immunological compatibility | C; D for the ABO, Rhesus, Kell, Kidd and Duffy markers listed in the rule | HLA typing for transplantation (C), ABO grouping kits (D) |
| 3 | Thirteen high-risk uses, from sexually transmitted agents to cancer, genetic and companion diagnostic tests | C | See Rule 3 in detail |
| 4 | Self-testing; near-patient testing | Self-tests C, with listed exceptions in B; near-patient tests classified in their own right | Glucose meters for self-testing (C), pregnancy self-tests (B) |
| 5 | General laboratory products, buffers, washing solutions, general culture media and stains made suitable for IVD procedures relating to a specific examination; instruments; specimen receptacles | A | Culture media with selective agents, NGS sequencers, blood spot cards |
| 6 | Everything not covered by another rule | B | Electrolyte tests, CRP tests, H. pylori and C. difficile tests, SARS-CoV-2 detection |
| 7 | Controls without a quantitative or qualitative assigned value | B | Unassigned control sera, multi-analyte control serum |
Read more: what “life-threatening” and “high risk of spreading” mean
MDCG 2020-16 describes life-threatening conditions as those that generally result in death, and are often untreatable, have limited treatment options or need major medical intervention.[3]
The risk of spreading is judged on how easily the agent is transmitted, contact rates and how long people are infectious, considering the general European population. MDCG 2020-16 notes that its list of high-risk agents may be updated as new evidence on incidence, pathogenicity and transmission emerges, so classification needs revisiting when the evidence changes.[3]
Rule 3 in detail: the thirteen Class C uses
Rule 3 catches most of the higher-risk tests that aren't in Class D. If Rule 1 or Rule 2 also applies, the higher class wins.[1][3]
| Point | Intended for | Examples |
|---|---|---|
| (a) | Detecting a sexually transmitted agent, or exposure to one | Chlamydia, gonorrhoea, HPV, syphilis |
| (b) | Detecting an infectious agent in cerebrospinal fluid or blood without a high or suspected high risk of spreading | Neisseria meningitidis, Cryptococcus neoformans, herpes simplex virus |
| (c) | Detecting an infectious agent where there is a significant risk that a wrong result would cause death or severe disability | MRSA, rubella, measles |
| (d) | Prenatal screening of immune status to transmissible agents | Rubella and toxoplasma immunity |
| (e) | Infective or immune status where a wrong result could lead to a life-threatening decision | Rubella IgM in pregnancy, interferon-gamma release assays |
| (f) | Companion diagnostics | ALK testing in non-small-cell lung cancer, PD-L1 testing |
| (g) | Disease staging where a wrong result could lead to a life-threatening decision | BNP for heart failure severity |
| (h) | Cancer screening, diagnosis or staging | Faecal immunochemical tests, CA 125 |
| (i) | Human genetic testing | BRCA1/2, factor V and prothrombin gene variants, pharmacogenomic tests |
| (j) | Monitoring levels where a wrong result could lead to a life-threatening decision | Troponin, lithium, tacrolimus |
| (k) | Managing patients with a life-threatening disease or condition | CD4 counts in HIV, INR, HbA1c |
| (l) | Screening embryos or fetuses for congenital disorders | Non-invasive prenatal testing for aneuploidy |
| (m) | Newborn screening where missed disorders could be life-threatening or disabling | Phenylketonuria, congenital hypothyroidism |
Self-testing and near-patient testing
Self-testing: Class C by default
A device intended to be used by lay persons, including testing services offered to lay persons online. Self-tests are Class C, except tests for pregnancy, fertility and cholesterol, and for glucose, erythrocytes, leucocytes and bacteria in urine, which are Class B.[1]
Higher rules still win
If Rule 1 or 2 applies, the self-test takes that higher class: MDCG 2020-16 lists an HIV self-test as Class D. Collecting your own sample doesn't make it a self-test; the lay person must do at least part of the test.[3]
Companion diagnostics
A companion diagnostic is a device essential for the safe and effective use of a corresponding medicinal product, to identify the patients most likely to benefit, or those at increased risk of serious adverse reactions, before or during treatment. Companion diagnostics are Class C under Rule 3(f).[1]
Before issuing a certificate, the notified body consults the European Medicines Agency or a national medicines authority on the suitability of the device for the medicinal product. MDCG 2020-16 notes that devices for monitoring treatment to keep drug levels within the therapeutic window, or for setting the dose for patients already eligible for the medicine, aren't companion diagnostics.[1][3]
Software, calibrators, controls and kits
Software
Software that only drives or influences a device takes that device's class. Software independent of any other device is classified in its own right: MDCG 2019-11 gives HbA1c software as Class C under Rule 3(k).[1][4]
Calibrators and assigned controls
Calibrators, and controls with quantitative or qualitative assigned values, take the same class as the device they're intended for.[1]
Unassigned controls
Controls without an assigned value are Class B under Rule 7, unless another rule applies. MDCG 2020-16 Rev.5 clarifies that concentration ranges a manufacturer indicates for standalone QC materials aren't assigned values. Internationally certified reference materials and materials for external quality assessment schemes are outside the IVDR altogether.[1][3]
Transition deadlines for legacy devices
Regulation (EU) 2024/1860 extended the IVDR transition periods, provided conditions are met. Devices with a valid certificate under the old IVD Directive can stay on the market until 31 December 2027, whatever their IVDR class, if the manufacturer applied to a notified body by 26 May 2025 and signed an agreement by 26 September 2025.[5]
Devices that were self-declared under the directive but need a notified body under the IVDR have dates by class:[5][6]
| IVDR class | Can stay on the market until | Apply to a notified body by | Signed agreement by |
|---|---|---|---|
| D | 31 December 2027 | 26 May 2025 | 26 September 2025 |
| C | 31 December 2028 | 26 May 2026 | 26 September 2026 |
| B and sterile A | 31 December 2029 | 26 May 2027 | 26 September 2027 |
Other conditions: the device keeps complying with the directive, has no significant changes in design or intended purpose and presents no unacceptable risk, and the manufacturer had an IVDR quality management system in place by 26 May 2025.[5][6]
What's new in the guidance
- MDCG 2020-16 Rev.4 (March 2025): SARS-CoV-2 was dropped from the Rule 1 examples. Detection and antibody tests now appear as Class B examples under Rule 6, and self-tests as Class C under Rule 4(a).[3]
- MDCG 2020-16 Rev.5 (September 2026): updates the rationale and examples for Rule 7 controls without assigned values. The classification rules themselves haven't changed.[3]
Common mistakes, and the fix
The same marker can be Class B or C depending on whether a wrong result could lead to a life-threatening decision. Start from the intended purpose.
Check Rules 1 to 3 even for self-tests and near-patient tests. The highest class applies.
Broad claims pull the device into higher classes. Be precise about the specimen, population and clinical use.
They're classified in their own right under the other rules.
With assigned values, they take the assay's class, not Class B.
Plan time for the notified body to consult the medicines authority.
A change in the spread or severity of an agent can change the class. Review it as part of post-market surveillance.
Record every rule you considered and why it does or doesn't apply. Notified bodies expect it.
Common questions
How many classification rules does the IVDR have?
Seven, in Annex VIII, plus implementing rules on how to apply them. They place each device in Class A, B, C or D.[1]
What is the default class under the IVDR?
Class B. Rule 6 says devices not covered by any other rule are Class B.[1]
Are all self-tests Class C?
No. Self-tests are Class C by default, but tests for pregnancy, fertility and cholesterol, and for glucose, erythrocytes, leucocytes and bacteria in urine are Class B. A self-test covered by Rule 1 or 2 is Class D.[1][3]
Do Class A IVDs need a notified body?
Not usually. Class A devices are self-declared, unless they're placed on the market sterile, in which case a notified body assesses the sterility aspects.[1]
What class is an IVD instrument?
Instruments specifically intended for IVD procedures are Class A under Rule 5(b). Instruments that measure directly through their own sensors without separate reagents, such as blood gas or glucose meters, are classified by their intended purpose, and glucose meters with strips for self-testing are Class C. Software that only drives the instrument takes the instrument's class.[1][3][4]
Is IVDR Class D the same as FDA Class III?
No. The EU and US systems use different rules. An IVD's class in one system doesn't decide its class in the other.
References
- Regulation (EU) 2017/746 on in vitro diagnostic medical devices (IVDR), in particular recitals 54 and 55, Article 1(3), Article 2 (definitions), Article 29, Article 47 (classification), Article 48 (conformity assessment), Articles 80 and 81, and Annex VIII (classification rules). eur-lex.europa.eu/eli/reg/2017/746/oj
- European Commission, Proposal for a Regulation amending Regulation (EU) 2017/746 as regards transitional provisions for certain in vitro diagnostic medical devices, COM(2021) 627, explanatory memorandum. data.consilium.europa.eu/doc/document/ST-12884-2021-INIT/en/pdf
- Medical Device Coordination Group, MDCG 2020-16 Rev.5, Guidance on classification rules for in vitro diagnostic medical devices under Regulation (EU) 2017/746 (September 2026), and its earlier revisions. health.ec.europa.eu/latest-updates/mdcg-2020-16-rev5-guidance-classification-rules-vitro-diagnostic-medical-devices-under-regulation-eu-2026-09-09_en
- Medical Device Coordination Group, MDCG 2019-11 Rev.1, Guidance on Qualification and Classification of Software in Regulation (EU) 2017/745 – MDR and Regulation (EU) 2017/746 – IVDR (June 2025). health.ec.europa.eu/document/download/b45335c5-1679-4c71-a91c-fc7a4d37f12b_en?filename=mdcg_2019_11_en.pdf
- Regulation (EU) 2024/1860 of 13 June 2024 amending Regulations (EU) 2017/745 and (EU) 2017/746 as regards a gradual roll-out of Eudamed, the obligation to inform in case of interruption or discontinuation of supply, and transitional provisions for certain in vitro diagnostic medical devices, in particular Article 110(3a) to (3c) IVDR as amended. eur-lex.europa.eu/eli/reg/2024/1860/oj
- Health Products Regulatory Authority (Ireland), IVDR transitional provisions. www.hpra.ie/regulation/medical-devices/manufacturers-and-authorised-representatives/ivdr-transitional-provisions
- European Commission, Proposal for a regulation to simplify rules on medical and in vitro diagnostic devices, COM(2025) 1023, procedure 2025/0404(COD), 16 December 2025. health.ec.europa.eu/publications/proposal-regulation-simplify-rules-medical-and-vitro-diagnostic-devices_en
This guide is for information only and isn't legal or regulatory advice. Classification depends on your intended purpose; confirm it against Annex VIII of the IVDR and the current MDCG guidance, and agree it with your notified body where one is involved.