Interactive guide · FDA · 11 min read
FDA Medical Device Classification: How to Find Your Device's Class, Product Code and Pathway
Every medical device sold in the US belongs to Class I, II or III, and that class decides whether you need a 510(k), a De Novo or a PMA, or no premarket submission at all. You don't work the class out from scratch: FDA has already classified about 1,700 generic types of device. This guide shows you how to find the one that matches yours, read its regulation, check whether it's exempt, and what to do when nothing fits.
Class I
General controls. Most are exempt from 510(k).
Class II
General and special controls. Usually a 510(k).
Class III
General controls and premarket approval (PMA).
How FDA classification works
Section 513 of the Federal Food, Drug, and Cosmetic Act sets up three classes based on the level of control needed to give reasonable assurance of a device's safety and effectiveness. FDA's procedures are in 21 CFR Part 860, and the individual classifications are in Parts 862 to 892.[1][2][3]
Already classified
FDA has classified about 1,700 generic types of device and grouped them into 16 medical specialty panels. Your job is usually to find the right one, not to create a class.[3][4]
Intended use decides
Classification depends on the device's intended use and its indications for use. Write those down before you search.[3]
Regulation number
Each classified type has a seven-digit regulation number, such as 21 CFR 880.2920 for the clinical mercury thermometer, which is Class II.[3]
Product code
Each type also has one or more three-letter product codes, such as FLK for that thermometer. You'll use the code in your submission and when you list the device with FDA.[3]
The three classes and what each one means for you
| Class | When a device goes in this class | Controls | Usual premarket route |
|---|---|---|---|
| I | General controls alone are enough for reasonable assurance of safety and effectiveness; or there isn't enough information to decide that or to set special controls, but the device isn't life-supporting or life-sustaining, isn't of substantial importance in preventing impairment of health, and doesn't present a potential unreasonable risk of illness or injury | General controls | Most are exempt from 510(k); the rest need one |
| II | General controls alone aren't enough, but there's enough information to set special controls | General controls and special controls | Usually a 510(k); some are exempt |
| III | There isn't enough information for general or special controls to assure safety and effectiveness, and the device is life-supporting or life-sustaining, of substantial importance in preventing impairment of health, or presents a potential unreasonable risk | General controls and premarket approval | PMA, unless it's a preamendment type for which FDA hasn't yet required PMAs |
Read more: general controls and special controls
General controls apply to every device, whatever its class. They come from the Act's provisions on adulteration, misbranding, registration, listing and premarket notification, banned devices, notification and other remedies (such as repair, replacement or refund), records, reports and unique device identification, and general provisions, which include good manufacturing practice, now the Quality Management System Regulation (QMSR).[2]
Special controls are extra requirements for a particular Class II device type, such as performance testing, labeling, postmarket surveillance or guidance. They're written into the device's classification regulation, so read it in full.[2]
How to find your device's classification, step by step
- Write down your intended use and indications
What the device does, for which condition and patients, where it's used, and whether it's prescription or over the counter. These decide the match.
- Search the Product Classification database
Search by part of the device name. FDA notes that the first word of a device name is usually the best search term. You can also browse by panel.[3][5]
- Shortlist the product codes that fit
Read each candidate's definition and regulation, not just the name. Names are short and can mislead.
- Read the classification regulation in the CFR
Each regulation has an identification paragraph, which describes the device type, and a classification paragraph, which gives the class, any special controls and any exemption.[6]
- Check the exemption and its limits
If the regulation says the device is exempt, check the limitations in section .9 of that part before relying on it. The checker below walks you through them.[7]
- Check what similar devices used
Search FDA's 510(k), De Novo and PMA databases for devices like yours and note the product codes they were cleared or approved under.
- Record your rationale
Note the codes you considered and why you chose one. If nothing fits, go to When nothing fits.
How to read a product classification record
Each product code has a record in FDA's database. These are the fields that matter most.[5]
| Field | What it tells you |
|---|---|
| Device and definition | The device type's name and, for many codes, a short definition of what's covered |
| Regulation number | The 21 CFR section that classifies the device. If there's no regulation number, the class isn't final. |
| Device class | 1, 2 or 3, or a note that it isn't classified, with the reason, such as preamendment or enforcement discretion |
| Submission type | What FDA expects before marketing: 510(k), PMA, 510(k) exempt, HDE, enforcement discretion or emergency use authorization, or contact FDA |
| Medical specialty, review panel and premarket review | The specialty the regulation sits under, the panel that reviews the device, and the FDA review division |
| GMP exempt | Whether the device is exempt from most QMSR requirements (records and complaint files under 21 CFR 820.35 still apply) |
| Third party review | Whether an accredited third party can review a 510(k) for it |
| Implant and life-sustaining flags | Whether FDA treats the type as an implant or as life-sustaining or life-supporting |
| Summary malfunction reporting | Whether the code is eligible for FDA's Voluntary Malfunction Summary Reporting Program (quarterly summaries instead of individual reports)[8] |
Do you need a 510(k)? The exemption checker
Use this once you've found a classification regulation that fits. It applies the exemption limits that FDA writes into section .9 of each part. Confirm the result against the regulation itself, and ask FDA through a 513(g) or a Pre-Submission if you're unsure.
1. Did you find a product code and classification regulation that fit your intended use?
2. What class is it?
3. What submission type does the record show?
3. Does the regulation say the device is exempt from premarket notification (510(k))?
4. Tick anything that's true of your device
Exemptions, and their limits
510(k) exemptions
FDA says about 572 Class I device types, around 74%, are exempt from premarket notification. Some Class II types are exempt too. Each exemption is written into the device's regulation, sometimes with conditions.[3]
Limits in section .9
An exemption only covers devices with the existing or reasonably foreseeable characteristics of commercially distributed devices of that type. A different intended use or a different fundamental scientific technology means you need a 510(k).[7]
GMP exemptions
Some exempt devices are also exempt from most of the QMSR. They still have to meet the requirements for records and complaint files in 21 CFR 820.35.[9]
Read more: the IVD uses that remove an exemption
For an IVD, the exemption doesn't apply if a misdiagnosis could carry high morbidity or mortality. Section .9(c) lists these uses:[7]
- Diagnosing, monitoring or screening for neoplastic diseases (cancers and other tumours), except immunohistochemical devices
- Screening or diagnosing familial or acquired genetic disorders, including inborn errors of metabolism
- Measuring a surrogate marker to screen for, diagnose or monitor life-threatening diseases such as AIDS, chronic or active hepatitis, tuberculosis or myocardial infarction, or to monitor therapy
- Assessing the risk of cardiovascular diseases
- Managing diabetes
- Identifying or inferring the identity of a microorganism directly from clinical material
- Detecting antibodies to microorganisms, except qualitative IgG assays on serum or plasma that aren't used to determine immunity
- Noninvasive testing
- Near-patient testing (point of care)
When nothing fits: unclassified and new devices
New device types are Class III by default
A device that wasn't on the market before 28 May 1976 is automatically in Class III unless it's substantially equivalent to a preamendment device or to a device already in Class I or II, or FDA classifies it otherwise.[1]
De Novo for low and moderate risk
If general controls, or general and special controls, would be enough, you can ask FDA to classify the device into Class I or II through a De Novo request. If granted, it creates a new classification regulation that later devices can use as a predicate.[10]
Unclassified product codes
Some product codes have no regulation, for reasons such as preamendment status or enforcement discretion. The record's submission type tells you what FDA expects.[5]
Is it a device at all?
Some products are drugs, biologics or combination products, or aren't regulated as devices. Our 510(k), De Novo or PMA guide covers how to check, and who reviews combination products.
Ask FDA: 513(g) requests for information
A 513(g) request asks FDA what class your device is in and which requirements apply. It's useful when you can't decide between product codes, or aren't sure your product is a device.[11]
| Question | Answer |
|---|---|
| What you get | FDA's view of the generic type and class, whether a 510(k), a PMA or neither is needed, and other requirements that apply |
| What you don't get | A classification decision, clearance or approval. FDA doesn't review substantial equivalence or safety and effectiveness data in a 513(g), and its response isn't final agency action. |
| How long | FDA intends to respond within 60 days of receipt |
| What to send | A cover letter with your questions, a device description, the intended use and proposed labeling. One product per request. |
| Fee | $8,596, or $4,298 for qualifying small businesses, for fiscal year 2027 (October 2026 to September 2027)[12] |
A Pre-Submission (Q-Submission) is the better route when you want FDA's feedback on testing or a submission plan rather than the class.[11]
Reclassification and accessories
Reclassification
FDA can move a device type to another class when new information shows a different level of control is right, and manufacturers can petition for postamendment Class III devices to be reclassified.[1]
Accessories classified on their own risk
Since FDARA amended section 513(f) in August 2017, FDA classifies an accessory based on its own risks when used as intended, regardless of the parent device's class. An accessory to a Class III device can be Class I or II.[1][13]
Accessory classification requests
You can ask FDA to classify a new accessory type alongside the parent device's 510(k) or PMA, or an existing one through a separate request. FDA must grant or deny an existing-accessory request within 85 days of receiving it.[13]
The 16 classification panels
Classification regulations are grouped by medical specialty. Some parts cover two specialties.[4]
| 21 CFR part | Part title |
|---|---|
| 862 | Clinical chemistry and clinical toxicology |
| 864 | Hematology and pathology |
| 866 | Immunology and microbiology |
| 868 | Anesthesiology |
| 870 | Cardiovascular |
| 872 | Dental |
| 874 | Ear, nose and throat |
| 876 | Gastroenterology-urology |
| 878 | General and plastic surgery |
| 880 | General hospital and personal use |
| 882 | Neurological |
| 884 | Obstetrical and gynecological |
| 886 | Ophthalmic |
| 888 | Orthopedic |
| 890 | Physical medicine |
| 892 | Radiology |
Selling in Europe too? The EU uses different rules and classes. See our guides to EU MDR classification and IVDR classification.
Common mistakes, and the fix
Read the identification paragraph of the regulation. Two codes with similar names can cover different uses.
If your indications differ, the code may not fit. Compare intended use, not just design.
Registration, listing, labeling, reporting and the QMSR still apply. GMP-exempt devices still keep records and complaint files under 21 CFR 820.35.
A new technology or use can remove an exemption. Check the limits every time.
Class II regulations often name specific testing or labeling. Build them into your plan early.
A new indication can move the device to another code or class, and may need a new submission.
It's the opposite: a new device type is Class III until FDA classifies it.
Record the codes you considered and why. You'll need it for submissions, audits and changes.
Common questions
How do I find my medical device's FDA classification?
Search FDA's Product Classification database by part of the device name or by panel, then read the classification regulation for each candidate. It tells you the class, any special controls and whether the device is exempt from 510(k).[3]
What is an FDA product code?
A three-letter code that identifies a generic device type. It links the device to its regulation number, class and review panel, and you use it in submissions and device listing.[3][5]
Do Class I devices need FDA approval?
Most don't need a premarket submission: FDA says about 74% of Class I device types are exempt from 510(k). They still follow general controls such as registration, listing and labeling.[3]
What happens if my device doesn't fit any classification?
A new type of device is automatically Class III. If its risk is low or moderate, you can ask FDA to classify it into Class I or II through a De Novo request.[1][10]
What is a 513(g) request?
A request asking FDA which class your device is in and which requirements apply. FDA intends to respond within 60 days, but the response isn't a classification decision or clearance.[11]
Is FDA Class II the same as EU Class IIa or IIb?
No. The FDA and EU systems use different rules, so a device's class in one doesn't decide its class in the other.
References
- Federal Food, Drug, and Cosmetic Act, section 513, Classification of devices intended for human use (21 U.S.C. 360c), including 513(a), 513(e), 513(f) and 513(g). www.govinfo.gov/link/uscode/21/360c
- 21 CFR Part 860, Medical Device Classification Procedures, in particular 860.3 (definitions of Class I, II and III, general controls and special controls). www.ecfr.gov/current/title-21/chapter-I/subchapter-H/part-860
- U.S. Food and Drug Administration, Classify Your Medical Device. www.fda.gov/medical-devices/overview-device-regulation/classify-your-medical-device
- U.S. Food and Drug Administration, Device Classification Panels. www.fda.gov/medical-devices/classify-your-medical-device/device-classification-panels
- U.S. Food and Drug Administration, Download Product Code Classification Files (field definitions for the Product Classification database). www.fda.gov/medical-devices/classify-your-medical-device/download-product-code-classification-files
- 21 CFR 880.2920, Clinical mercury thermometer (an example of a classification regulation's identification and classification paragraphs). www.ecfr.gov/current/title-21/chapter-I/subchapter-H/part-880/subpart-C/section-880.2920
- 21 CFR 862.9, Limitations of exemptions from section 510(k) of the Federal Food, Drug, and Cosmetic Act; the same limitations appear in section .9 of each of Parts 864 to 892. www.ecfr.gov/current/title-21/chapter-I/subchapter-H/part-862/subpart-A/section-862.9
- U.S. Food and Drug Administration, Voluntary Malfunction Summary Reporting Program. www.fda.gov/medical-devices/medical-device-reporting-mdr-how-report-medical-device-problems/voluntary-malfunction-summary-reporting-program
- U.S. Food and Drug Administration, Medical Devices; Quality Management System Regulation Technical Amendments, Federal Register, 90 FR 55978, 4 December 2025 (FR Doc. 2025-21955). www.federalregister.gov/documents/2025/12/04/2025-21955/medical-devices-quality-management-system-regulation-technical-amendments
- U.S. Food and Drug Administration, De Novo Classification Request. www.fda.gov/medical-devices/premarket-submissions/de-novo-classification-request
- U.S. Food and Drug Administration, FDA and Industry Procedures for Section 513(g) Requests for Information under the Federal Food, Drug, and Cosmetic Act, guidance (August 2024). www.fda.gov/regulatory-information/search-fda-guidance-documents/fda-and-industry-procedures-section-513g-requests-information-under-federal-food-drug-and-cosmetic
- U.S. Food and Drug Administration, Medical Device User Fee Rates for Fiscal Year 2027, Federal Register, 30 July 2026 (FR Doc. 2026-15335). www.federalregister.gov/documents/2026/07/30/2026-15335/medical-device-user-fee-rates-for-fiscal-year-2027
- U.S. Food and Drug Administration, Medical Device Accessories. www.fda.gov/medical-devices/classify-your-medical-device/medical-device-accessories
- U.S. Food and Drug Administration, Medical Devices; Classification of Accessories Distinct from Other Devices; Proposed List of Accessories Suitable for Class I; Request for Comments, Federal Register, 17 August 2026 (FR Doc. 2026-16729, docket FDA-2025-N-6224). www.federalregister.gov/d/2026-16729
This guide is for information only and isn't legal or regulatory advice. Always confirm your device's classification against the current classification regulation and FDA's database, and ask FDA when you're unsure.