Regulatory Update United States Regulatory Quality Manufacturing & Supply Chain IVD

US regulatory update · FDA (CDRH and CBER) · Final guidance issued 20 November 2026

FDA Finalizes Its 510(k) Clearance Transfer Q&A Guidance: What New and Previous Holders Need to Do

Why it matters

Acquisitions, divestitures and product-line sales often move 510(k) clearances between companies. On 19 November 2026 FDA posted the final guidance "Transfer of a Premarket Notification (510(k)) Clearance – Questions and Answers", dated 20 November 2026, with the Federal Register notice of availability scheduled for publication the same day [1][2]. It finalizes the June 2025 draft [1][3].

The guidance is non-binding, but most of its answers restate binding requirements in 21 CFR Parts 807, 801 and 830. The practical message is that a transfer does not by itself need a new 510(k), while registration, listing, labeling and GUDID records must still be brought into line, and those obligations sit with the companies, not with FDA [1].

What changed

Status: NEW final guidance (finalizes the June 2025 draft; docket FDA-2024-D-1528) [1][2]. The final document answers 12 questions. The main points, with the cited regulations, are [1]:

  • No new 510(k) for an unchanged device. If the device is not significantly changed or modified in design, components, method of manufacture or intended use, the new holder does not need to submit a new 510(k) (Q4). A significant change generally requires a new 510(k) under 21 CFR 807.81(a)(3) (Q7).
  • List with the original K-number. The new holder must list the device using the original FDA-assigned premarket submission number, unless it obtains a new 510(k) (21 CFR 807.25(g)(4); Q7). Listing generates a new listing number (Q5).
  • Register within 30 days. A new holder not previously engaged in an operation described in 21 CFR 807.20(a) must register within 30 days and submit listing information at that time (21 CFR 807.22(a); Q6).
  • One holder at a time. FDA considers there to be a single 510(k) holder for a device at a time (Q8). The previous holder must update its listing when it stops an activity, and discontinue the listing when it stops commercial distribution (21 CFR 807.28; Q8).
  • UDI and labels. A new listing number triggers a GUDID update (21 CFR 830.330(b)). Labeler information that changes must be updated in GUDID by the date a device is first labeled with it, or within 10 business days if it is not on the label. Labels may also need a new name and place of business (21 CFR 801.1(a); Q10).
  • FDA will not fix your records. FDA is not required to update registration or listing data in response to transfer notifications. That responsibility sits with the owner or operator (Q11). Failure to register, list, update GUDID or submit a required 510(k) may make the device adulterated and/or misbranded (Q10).

Note that the database "applicant" name does not change when a 510(k) is transferred (footnote 17) [1]. Contract manufacturers, contract sterilizers, repackagers and relabelers must list using the holder's K-number; initial importers may list by giving the manufacturer's name and address (Q9) [1].

After acquiring a 510(k) clearance: decision flow [1]

Is the device significantly changed or modified in design, components, method of manufacture or intended use?
YesNo
Yes →Yes: submit a new 510(k) (21 CFR 807.81(a)(3)); list with the new K-number after clearance (Q7)
No →No: no new 510(k) needed (Q4); list with the ORIGINAL FDA-assigned K-number (21 CFR 807.25(g)(4))

In both cases:

New holder: register within 30 days if not previously in an 807.20(a) operation; submit listing at that time (Q6)
New listing number: update GUDID record (21 CFR 830.330(b)); revise labels and GUDID labeler data as needed (Q10)
Previous holder: update the listing if it stops an activity; discontinue it if it stops commercial distribution (Q8)

One 510(k) holder per device at a time (Q8). FDA does not update registration or listing data on request; the owner or operator is responsible (Q11).

Failure to register, list, update GUDID or submit a required 510(k) may render the device adulterated and/or misbranded (Q10).

Figure 1. Post-transfer decision flow, summarised from the final guidance [1].

Who is affected

Medical devices

All 510(k)-cleared devices regulated by CDRH and CBER, including both parties to a sale or transfer, and the contract manufacturers, sterilizers, repackagers, relabelers and importers who list against that clearance [1].

IVDs

The guidance does not carve out IVDs. Interpretation it applies equally to IVDs cleared through 510(k), including CBER-regulated devices such as blood-establishment tests cleared through 510(k). IVDs authorised by De Novo or PMA are outside its scope. Any assay change made after the transfer (reagents, manufacturing site or method, or claims) still needs a separate analysis of whether a new 510(k) is required.

This is a US FDA guidance. It does not change EU MDR/IVDR rules on changing manufacturers or transferring certificates, which follow their own requirements.

Key dates

Table 1. Key dates

DateEventSource
22 Dec 2014First draft guidance announced[4]
20 Mar 20202014 draft withdrawn[5]
5 Jun 2025New draft guidance issued[3]
4 Aug 2025Comment target date for the 2025 draft[3]
19 Nov 2026Final guidance posted; FR notice placed on public inspection[1][2]
20 Nov 2026Final guidance issued; FR notice of availability scheduled for publication[1][2]

History of the 510(k) transfer Q&A guidance [1]–[5]

  1. 22 Dec 2014First draft guidance issued
  2. 20 Mar 20202014 draft withdrawn
  3. 5 Jun 2025New draft guidance issued
  4. 4 Aug 2025Draft comment target date
  5. 19 Nov 2026Posted on FDA site; FR notice on public inspection
  6. 20 Nov 2026Final guidance issued; FR publication
Figure 2. Timeline of the guidance [1]–[5].

The guidance does not state a separate implementation date. The 30-day registration period runs from entering into an 807.20(a) operation, under existing regulations [1].

What RA teams should do

  • Add a regulatory workstream to deal checklists. Cover registration, listing with the original K-number, GUDID updates, label updates and discontinuing the previous holder's listing, with owners and dates on both sides.
  • Run a change assessment before closing. Interpretation use FDA's "Deciding When to Submit a 510(k)" guidances (device and software) to decide whether any planned site, supplier, process or labeling change triggers a new 510(k) [1].
  • Keep proof of ownership. Interpretation the guidance relies on registration and listing rather than an FDA "transfer letter", so keep transaction documents and the full 510(k) file. If records are missing, ask the prior owner for copies or file a FOIA request (Q12) [1].
  • Line up supply-chain listings. Make sure contract manufacturers, sterilizers, repackagers and relabelers list against the correct K-number (Q9) [1].
  • Update procedures. Check registration and listing, UDI and labeling SOPs against the final Q&A, and record the guidance in the regulatory requirements register.

References

  1. FDA (CDRH/CBER). Transfer of a Premarket Notification (510(k)) Clearance – Questions and Answers: Guidance for Industry and Food and Drug Administration Staff. Document issued 20 November 2026; webpage content current as of 19 November 2026. https://www.fda.gov/regulatory-information/search-fda-guidance-documents/transfer-premarket-notification-510k-clearance-questions-and-answers (PDF: https://www.fda.gov/media/186765/download). Verified 19 November 2026.
  2. Office of the Federal Register. Guidance: Transfer of a Premarket Notification (510(k)) Clearance, Questions and Answers. FR Doc. 2026-23668, filed for public inspection 19 November 2026, scheduled publication 20 November 2026. https://www.federalregister.gov/d/2026-23668. Verified 19 November 2026.
  3. FDA. Transfer of a Premarket Notification (510(k)) Clearance—Questions and Answers; Draft Guidance; Availability. Federal Register, 5 June 2025 (FR Doc. 2025-10238). https://www.federalregister.gov/documents/full_text/html/2025/06/05/2025-10238.html. Verified 19 November 2026.
  4. FDA. Transfer of a Premarket Notification (510(k)) Clearance—Questions and Answers; Draft Guidance; Availability. Federal Register, 22 December 2014 (79 FR 76331). https://www.federalregister.gov/documents/2014/12/22/2014-29832/transfer-of-a-premarket-notification-510k-clearance-questions-and-answers-draft-guidance-for. Verified 19 November 2026.
  5. FDA. Withdrawn or Expired Guidance (Medical Devices). https://www.fda.gov/medical-devices/guidance-documents-medical-devices-and-radiation-emitting-products/withdrawn-or-expired-guidance. Verified 19 November 2026.

Disclaimer: This post is regulatory intelligence for information only and is not legal or regulatory advice. Statements labelled INTERPRETATION are the author's analysis. Confirm requirements against the primary sources before acting.

Previous
Previous

How to Create and Maintain an Approved Supplier List (ASL)

Next
Next

Planned and Unplanned Deviations: What They Are and How to Handle Them