UDI in the US and EU: GUDID vs EUDAMED, Basic UDI-DI, Deadlines and What's Different

A unique device identifier (UDI) is the code on a medical device label that lets anyone in the supply chain identify the exact device and, usually, its lot or serial number. The US and the EU both require one, built on the same international idea, but the rules differ: different databases, different deadlines, and an extra identifier in the EU called the Basic UDI-DI. This guide compares the two side by side and includes a picker that shows what applies to your device.

UDI-DI

Device identifier: the fixed part that identifies the model and version

UDI-PI

Production identifier: lot or serial number, dates, as they apply

Basic UDI-DI

EU only: identifies a device model in EUDAMED and on certificates

GUDID / EUDAMED

Where device data goes in the US and the EU

What a UDI is

Two parts

The device identifier (UDI-DI in EU terms) identifies the labeler or manufacturer and the specific version or model. The production identifier (UDI-PI) adds the lot or batch, serial number, expiration date or manufacturing date, as they apply. US regulations simply say device identifier and production identifier.[1][2]

Two forms

The UDI appears in plain text people can read and in a form machines can read, such as a linear or 2D barcode.[3][2]

Issued by accredited systems

The codes follow the standards of an issuing agency (US) or issuing entity (EU). GS1, HIBCC and ICCBBA are accredited in the US; the EU designated those three and IFA.[4][5]

Every packaging level

Each level of packaging, such as a unit and a box of units, gets its own UDI-DI. Shipping containers are excluded.[1][2]

US and EU, side by side

United StatesEuropean Union
Legal basis21 CFR Part 801 Subpart B (labeling) and Part 830 (UDI system and GUDID)MDR Article 27 and Annex VI Part C; IVDR Article 24 and Annex VI Part C
Who assigns itThe labelerThe manufacturer
DatabaseGlobal Unique Device Identification Database (GUDID)EUDAMED, UDI and device registration module
Model-level identifierNoneBasic UDI-DI, used in EUDAMED, certificates, declarations of conformity and technical documentation
IssuersFDA-accredited issuing agencies: GS1, HIBCC, ICCBBADesignated issuing entities: GS1, HIBCC, ICCBBA, IFA
Reusable devicesDirect marking for devices meant to be used more than once and reprocessed between uses, with exceptionsDirect marking for reusable devices, with exceptions
Dates on labelsYear-month-day format (for example, 2027-01-31)No set format. The use-by date must show at least year and month (MDR Annex I 23.2; IVDR Annex I 20.2, in year-month-day order). The human-readable UDI follows the issuing entity's rules.
Not coveredExceptions in 21 CFR 801.30, such as custom devices and investigational devicesCustom-made devices and devices for clinical investigation or performance studies
Quality recordsThe UDI is recorded for each device or batch (21 CFR 820.35(c))The manufacturer keeps a list of all assigned UDIs in the technical documentation. Under the MDR, economic operators and health institutions store the UDIs of Class III implantable devices they supply or receive; under the IVDR, only devices the Commission names

Summarized from 21 CFR Parts 801, 820 and 830, the MDR and the IVDR.[3][1][2][6][7]

What applies to your device? The picker

1. Which market?

The Basic UDI-DI: the EU's extra identifier

The Basic UDI-DI is the main identifier of a device model in the EU. It's assigned at the level of the device unit of use, it's the main key for records in EUDAMED, and it's referenced in certificates, the EU declaration of conformity, the technical documentation and, where required, the summary of safety and (clinical) performance. It doesn't appear on the label.[2][8]

One Basic UDI-DI, many UDI-DIs

Devices with the same intended purpose, risk class and essential design and manufacturing characteristics can share a Basic UDI-DI, while each version, size and packaging level has its own UDI-DI.[8][9]

Plan it early

Because it links certificates, declarations and EUDAMED records, changing how you group devices later means updating all of them.

EUDAMED device registration

  1. 27 November 2025

    The Commission published its decision confirming that four EUDAMED modules, including UDI and device registration, are functional.[10]

  2. 28 May 2026

    Using those modules became mandatory. New devices must be registered before the first unit is placed on the EU market.[11][12]

  3. By 28 November 2026

    Devices placed on the market before 28 May 2026, and still being placed on the market after it, must be registered within 12 months of the notice's publication, by 28 November 2026.[12][11]

Legacy devices (still sold under a valid directive certificate or declaration) must also be registered within the same window, unless the same device is already registered as an MDR or IVDR device. "Old" devices, placed on the market only before the MDR or IVDR applied, can't be registered. Devices no longer placed on the market only need registering if a vigilance or post-market surveillance action is needed.[12][11]

Labeling and direct marking deadlines

DeviceUDI on labelsDirect marking (reusable devices)
US: implantable, life-supporting or life-sustaining24 September 201524 September 2015 (implants don't need direct marking)
US: Class III24 September 201424 September 2016
US: Class II24 September 201624 September 2018
US: Class I and unclassified24 September 2018 (FDA enforcement from 24 September 2022)24 September 2020 (FDA enforcement from 24 September 2022)
EU MDR: Class III and implantable26 May 202126 May 2023
EU MDR: Class IIa and IIb26 May 202326 May 2025
EU MDR: Class I26 May 202526 May 2027
EU IVDR: Class D26 May 2023No separate date (applies with labels)
EU IVDR: Class B and C26 May 2025No separate date (applies with labels)
EU IVDR: Class A26 May 2027No separate date (applies with labels)

US dates from FDA's compliance dates page, including its enforcement policy for Class I and unclassified devices; EU dates from the Commission's UDI helpdesk.[13][14]

When you need a new UDI-DI

In both systems, a change that creates a new version or model of the device, or a new packaging configuration, needs a new UDI-DI. In the EU, Annex VI lists triggers such as a change to the name or trade name, device version or model, labeled as single use, packaged sterile, need for sterilization before use, quantity in a package, and critical warnings or contraindications (such as latex or DEHP). Commission guidance (MDCG 2018-1) adds a change in CMR or endocrine-disrupting substance labeling.[1][2][9]

Software in the EU

A new UDI-DI is needed when a change modifies the original performance, safety or intended use, or the interpretation of data. Minor revisions, such as bug fixes and security patches, need a new UDI-PI instead.[2]

Route it through change control

Build a UDI check into your change process, so labels, GUDID and EUDAMED are updated together. See our design change checklist.

Common mistakes, and the fix

Treating GUDID and EUDAMED as the same.

They need different data, and only EUDAMED uses the Basic UDI-DI. Map your data to both.

Grouping models under one Basic UDI-DI that don't belong together.

Check intended purpose, risk class and essential design and manufacturing characteristics match.

Changing a device without checking the UDI-DI.

Use the triggers as a checklist in every change.

Missing packaging levels.

Each level of packaging needs its own UDI-DI.

Forgetting reusable devices.

Plan direct marking that survives your reprocessing method.

Leaving EUDAMED to the last minute.

Devices already on the market had to be registered by 28 November 2026, and every new device must be registered before it's placed on the market. Data cleanup takes longer than you think.

Common questions

What is the difference between UDI-DI and Basic UDI-DI?

The UDI-DI is on the label and identifies a specific version and packaging level. The Basic UDI-DI is an EU-only identifier for the device model, used in EUDAMED, certificates and declarations of conformity, and not on the label.[2]

Is GUDID data accepted in EUDAMED?

No. They're separate databases with different data requirements, so you register in each.[1][2]

When did EUDAMED UDI registration become mandatory?

On 28 May 2026, six months after the Commission's notice was published in the Official Journal (27 November 2025). Devices already on the market had to be registered by 28 November 2026.[11][12][10]

Do Class I devices need a UDI?

Generally, yes. In the US, Class I devices need a UDI unless an exception applies, such as Class I devices FDA has exempted from good manufacturing practice requirements (21 CFR 801.30(a)(2)); a UPC on a Class I device serves as its UDI (801.40(d)). FDA began enforcing for Class I and unclassified devices in 2022. In the EU, Class I UDI labels have been required since 26 May 2025, and direct marking of reusable Class I devices from 26 May 2027.[13][14]

Which issuing agencies can I use?

In the US, the FDA-accredited issuing agencies GS1, HIBCC and ICCBBA. In the EU, the designated issuing entities GS1, HIBCC, ICCBBA and IFA.[4][5]

References

  1. 21 CFR Part 830, Unique Device Identification, including the definitions in § 830.3 and the GUDID requirements in Subpart E. www.ecfr.gov/current/title-21/chapter-I/subchapter-H/part-830
  2. Regulation (EU) 2017/745 on medical devices (MDR), Articles 27 to 29, Annex I section 23.2 and Annex VI Part C. eur-lex.europa.eu/eli/reg/2017/745/oj
  3. 21 CFR Part 801, Subpart B, Labeling requirements for unique device identification, including §§ 801.20, 801.30, 801.40 and 801.45, and Subpart A § 801.18 (format of dates). www.ecfr.gov/current/title-21/chapter-I/subchapter-H/part-801
  4. U.S. Food and Drug Administration, Contact an FDA-Accredited Issuing Agency. www.fda.gov/medical-devices/unique-device-identification-system-udi-system/contact-fda-accredited-issuing-agency
  5. Commission Implementing Decision (EU) 2019/939 of 6 June 2019 designating issuing entities designated to operate a system for the assignment of Unique Device Identifiers (UDIs) in the field of medical devices. eur-lex.europa.eu/eli/dec_impl/2019/939/oj
  6. Regulation (EU) 2017/746 on in vitro diagnostic medical devices (IVDR), Articles 24 to 26, Annex I section 20.2 and Annex VI Part C. eur-lex.europa.eu/eli/reg/2017/746/oj
  7. 21 CFR 820.35(c), Quality Management System Regulation, effective 2 February 2026. www.ecfr.gov/current/title-21/chapter-I/subchapter-H/part-820
  8. European Commission, UDI Helpdesk, UDI assignment: identifiers. webgate.ec.europa.eu/udi-helpdesk/en/udi-assignment/identifiers.html
  9. Medical Device Coordination Group, MDCG 2018-1, Guidance on basic UDI-DI and changes to UDI-DI, as revised. health.ec.europa.eu/medical-devices-sector/new-regulations/guidance-mdcg-endorsed-documents-and-other-guidance_en
  10. Commission Decision (EU) 2025/2371 of 26 November 2025 on the notice regarding the functionality and the fulfilment of the functional specifications of certain electronic systems included in the European Database on Medical Devices referred to in Article 34(1) of Regulation (EU) 2017/745, OJ L, 27.11.2025. eur-lex.europa.eu/eli/dec/2025/2371/oj
  11. Health Products Regulatory Authority (Ireland), Mandatory use of EUDAMED modules for medical devices. www.hpra.ie/news-events/news/article/mandatory-use-of-eudamed-modules-for-medical-devices
  12. European Commission, Questions and answers on the gradual roll-out of Eudamed under Regulation (EU) 2024/1860, Q7, Q8 and Q14. health.ec.europa.eu/document/download/0e7327c7-0e06-4fbd-90d3-8ab7bb30fe9f_en?filename=md_eudamed-qa_en.pdf
  13. U.S. Food and Drug Administration, Compliance Dates for UDI Requirements. www.fda.gov/medical-devices/unique-device-identification-system-udi-system/compliance-dates-udi-requirements
  14. European Commission, UDI Helpdesk, UDI carrier: deadlines. webgate.ec.europa.eu/udi-helpdesk/en/udi-carrier/deadlines.html