Medical Device Risk Management Tools
Hazard analysis, FMEA, use-related risk analysis, cybersecurity: each risk tool is a different lens on the same ISO 14971 process.[1] This guide shows how each one works, how to score risk the way ISO 14971 intends, and how the results come together in one risk management file.
What do you need to do?
One device runs through every part: the Orvelo, a fictional wearable ECG patch with a phone app and an AI arrhythmia-detection algorithm. Meet the device →
Part 1 of 11 · Field guide · 7 min read
One Process, Many Tools: The ISO 14971 Framework
Start here: the vocabulary every risk tool shares, how risk is scored, and which tool to reach for at each stage.
Meet the device
Every part of this guide uses the same fictional product, the Orvelo patch, invented for this guide. Any resemblance to a real device is unintended. Using one device throughout means you can see how the tools connect rather than learning each one in isolation.
Orvelo wearable ECG patch
A single-use adhesive patch that records a patient's ECG continuously for up to seven days. It runs on a non-rechargeable coin-cell battery.
HardwarePhone app and cloud
The patch streams data over Bluetooth to a phone app, which uploads it to a cloud service. Patients apply the patch themselves at home.
Software · connectivityAI arrhythmia detection
A machine learning model flags possible atrial fibrillation episodes for a clinician to review. It doesn't diagnose on its own.
Machine learningThat mix of skin contact, a battery, lay users, wireless data and an algorithm is why one device needs nearly every tool in this guide.
One process, many tools
ISO 14971:2019 is the international standard for medical device risk management, and regulators in the US, EU and most other markets expect it.[1] It defines a single process. The tools in this guide are methods for carrying out parts of it, and none of them is a risk management process on its own. A DFMEA spreadsheet, however thorough, is not a risk management file.
Some hazard areas also have their own specialist evaluations, such as biological safety under ISO 10993-1 and electrical and mechanical safety under IEC 60601-1. Their results feed the same risk management file. This guide focuses on the tools that tie everything together.[3]
Try it
The ISO 14971 process, and where each tool fits
Where the requirement lives
ISO 14971:2019 sets the process: plan, analyse, evaluate, control, review overall residual risk, then keep monitoring after release. Its companion ISO/TR 24971:2020 gives guidance and describes the analysis techniques.[1][4]
FDA QMSR (21 CFR 820) incorporates ISO 13485:2016, which requires documented risk management processes in product realization and points to ISO 14971 for how to do it.[5]
EU MDR and IVDR, Annex I require manufacturers to reduce risks as far as possible without adversely affecting the benefit-risk ratio, and to run a risk management system across the device's lifecycle.[6][7]
Hazard, hazardous situation, harm
ISO 14971 separates three ideas that everyday speech blurs together. A hazard is a potential source of harm, such as heat, electricity or a wrong number on a screen. A hazardous situation is the circumstance in which someone is exposed to it. Harm is the injury or damage that can follow. Risk is the combination of how severe that harm is and how likely it is to happen.[1]
Between a hazard and a harm is a sequence of events. Two probabilities matter along that chain: P1, the chance the hazardous situation occurs, and P2, the chance it then leads to harm. ISO 14971 illustrates this chain in its Annex C.[1] Pick an Orvelo example to see it.
Try it
From hazard to harm on the Orvelo
Quick quiz
Hazard, hazardous situation or harm?
Statement 1 of 6
Scoring risk: severity, probability and your policy
To evaluate a risk you need two scales, severity and probability, and a rule for which combinations are acceptable. ISO 14971 doesn't supply those criteria. It requires top management to set a policy for establishing them, and the risk management plan to define them for each product.[1] The matrix below uses example scales. Click any cell to change its zone: your edits carry through to every matrix in this guide.
Try it
Your risk acceptability policy
| Severity | Example definition | Probability | Example range (per use) |
|---|
A green cell isn't the end of the job in the EU
The EU MDR asks you to reduce risks as far as possible, not just below an acceptability line. A risk in your "acceptable" zone may still need further reduction if a further control is possible that doesn't worsen the benefit-risk balance. Under the prevailing EU interpretation, which comes from the harmonized standard's annexes and notified body practice rather than the MDR's wording, cost isn't an accepted reason to stop; the limits are the state of the art and not worsening the benefit-risk ratio. Part 10 explains how this differs from ALARP.[6]
Which tool, when?
Most tools are used at more than one stage, but each one is strongest at a particular point. Pick where you are to see the tools that apply.
Tool finder
Where are you in the lifecycle?
Part 2 of 11 · Hazard analysis · 7 min read
Hazard Analysis: From Hazard to Harm, With Numbers
Hazard analysis is the backbone of the risk management file. Every other tool in this guide feeds new hazardous situations into it, or estimates the probabilities that sit on its chains.
What hazard analysis is for
A hazard analysis works from the top down. Instead of asking how each part can fail, it asks what could hurt someone and how it could happen. Early in a project it's often called a preliminary hazard analysis (PHA): a first list of hazards drawn from the intended use, similar devices and known problems, made before the design is detailed. As the design matures, the same document grows into the full hazard analysis, with probabilities, controls and verification evidence on every chain.[4]
The most reliable way to build the list is to walk through hazard categories systematically, rather than brainstorming freely.
Prompt cards
Hazard categories, applied to the Orvelo
Estimating risk with P1 × P2
For each chain, multiply the probability the hazardous situation occurs (P1) by the probability it leads to harm (P2). The product is the probability of harm, which is what goes on the probability axis. Then add risk controls and watch where each one acts: some lower P1, others lower P2.[1]
Try it
P1 × P2 calculator: battery overheating
No data yet? Use levels instead of numbers
ISO 14971 doesn't require numerical probabilities. Early in design, most teams estimate P1 and P2 as qualitative or semi-quantitative levels, and switch to numbers where they have field, test or reliability data.[1][4] Switch the calculator to Levels to see how it works: you can't multiply words, so the plan defines a combination table that turns each P1 and P2 pair into one probability level for the matrix.
Two rules keep it honest. Define every level in words, so two engineers would pick the same one. And don't multiply rank numbers (P1 = 2 times P2 = 3); like the RPN in Part 3, the result looks like math but doesn't behave like a probability. Where probability can't be estimated at all, as with software failures or cyberattacks, ISO 14971 allows the evaluation to rest on severity, and security uses exploitability instead (Part 6, Part 7).
Notice how little the warning label changes the result. That's not a quirk of the example: information for safety is the weakest kind of control, which is why ISO 14971 puts it last in the order of preference. More on that in Part 9.
The Orvelo hazard analysis, at a glance
Here are five chains from the Orvelo analysis, scored against your policy from Part 1. Edit your policy and this table updates. Notice that the top two rows by severity aren't the top two by probability: you need both to decide where to spend effort.
| Hazardous situation | Harm | Sev. | P1 | P2 | P(harm) | Zone |
|---|
Getting the numbers right
Field complaint rates for similar devices, published adverse event data, component reliability data and your own test results all beat a guess. Record the source of every probability so a reviewer can follow your reasoning.
Some probabilities can't be estimated meaningfully, such as a software fault or a deliberate attack. In those cases, be explicit, rate the probability conservatively and let severity drive the decision about controls. Parts 6 and 7 cover how.
Score the severity of the actual harm at the end of the chain. "Electrical energy" has no severity; a superficial burn and a cardiac arrest do, and they belong on separate chains.
The same hazard can reach different harms through different sequences. Low battery can mean a data gap (missed diagnosis) on one chain and nothing worse than an annoyance on another. List them separately.
Part 3 of 11 · DFMEA · 7 min read
Design FMEA, and Why the Risk Priority Number Misleads
A DFMEA works from the bottom up: how can each part or function of the design fail, and what happens if it does? It's the most widely used risk tool in MedTech, and the most often misused.
How a DFMEA works
Failure mode and effects analysis goes component by component, or function by function. For each one you list how it could fail (the failure mode), what that failure would cause (the effect) and why it could happen (the cause). The standard for the method is IEC 60812:2018, and ISO/TR 24971 lists FMEA among the techniques that support ISO 14971 risk analysis.[8][4]
Each row usually gets three ratings, often on 1–10 scales: severity of the effect, occurrence of the cause, and detection, meaning how likely your design controls are to catch the problem before the design is released. Multiplying them gives the risk priority number, or RPN.
Two different questions
An FMEA asks how the design can fail. A hazard analysis asks how the patient or user can be harmed. A failure mode with no path to harm still matters for reliability, but it doesn't belong in your risk evaluation. A harm with no failure mode behind it (a use error, say, or a correct device used in the wrong patient) won't show up in a DFMEA at all. You need both views.
The trouble with RPN
ISO 14971 defines risk as the combination of the severity of harm and the probability of that harm occurring. Detection isn't part of it.[1] In a DFMEA, a good detection rating means your verification testing is likely to catch the problem in development. That's valuable, but it doesn't make the released device any safer: if the problem is caught, you fix the design, and then the occurrence rating should go down. Counting detection again in the RPN rewards the same work twice, and lets a severe failure mode hide behind a high detection score.
The automotive industry reached a similar conclusion. The AIAG-VDA FMEA Handbook of 2019 replaced RPN with an "action priority" approach that gives severity more weight.[9] Try it yourself: edit the ratings and compare the two rankings.
Try it
Orvelo DFMEA: rank by RPN, or by severity and occurrence?
| Row | Failure mode → effect | S | O | D | RPN |
|---|
How the rankings differ
With the starting ratings, the two failure modes that could seriously injure a patient, a battery short and a DC current path through the electrodes, sit at the bottom of the RPN list because they're rare and easy to detect in testing. Ranked by severity first, they rise to the top, which is where a regulator would expect to see your attention.
Linking the DFMEA to the hazard analysis
The hazard analysis works top-down, from hazard to hazardous situation to harm. The DFMEA works bottom-up, from a cause in one component to its effect on the device. They meet where a DFMEA row's end effect is a hazardous situation. Link the row there, and the failure mode and its cause become part of the sequence of events that leads to that hazardous situation.[1][4]
Notice that the DFMEA above uses a 1–10 severity scale while the hazard analysis in Part 2 uses five levels. That is common, and it's exactly how the two documents end up disagreeing. Once a row is linked, its severity should come from the harm in a shared harms table, not be scored again.
Try it
Trace each DFMEA row to the hazard analysis
Pick the end effect from a shared list
Keep one list of hazardous situations, each with an ID, and one harms table that sets each harm's severity once. The DFMEA's end-effect column picks from that list instead of free text.
Expect many-to-many links
One hazardous situation usually has several causes from the DFMEA, the PFMEA, the URRA and software. One failure mode can also lead to more than one hazardous situation.
Record "no safety impact" as a decision
Rows with no safety-related effect stay in the DFMEA as reliability or quality items. Mark them explicitly, with a short rationale, so a reviewer can tell the row was judged rather than forgotten.
Carry occurrence across, not detection
A cause's occurrence feeds P1 for the hazardous situation. Detection by testing or inspection doesn't lower patient risk, so it stays a verification question. Detection by the device itself, such as an alarm, is different: it's a protective measure and belongs in the hazard analysis as a risk control.
Keep one record per risk control
If current limiting appears in both the DFMEA and the hazard analysis, it should be the same control with one verification link. Two copies drift apart, and then there are two verification stories for one control.
Part 4 of 11 · PFMEA · 6 min read
Process FMEA and the Control Plan
A good design can still be built badly. A PFMEA asks how each manufacturing step can fail to produce what the design requires, and the control plan turns its answers into daily checks on the line.
Walk the process, step by step
A PFMEA follows the process flow. For each step it lists how the step can go wrong, what that does to the product, why it might happen, and the controls that prevent or detect it. The severity of each effect should come from the DFMEA and hazard analysis, so the whole risk file uses one view of how bad an outcome is. Click an Orvelo process step.
Try it
Orvelo process flow and PFMEA
Prevention, detection and what reaches the patient
Here the detection column means something different from the DFMEA. In production, an inspection that catches a bad unit before it ships really does lower how often a defect reaches a patient, and ISO 14971 recognizes protective measures in the manufacturing process as a form of risk control.[1] But detection has to be validated, and it is never perfect. Prevention stops the defect being made at all. Try both.
Try it
Escapes calculator: weak laminations
The escape rate is a direct input to P1 in the hazard analysis. If the production line changes, so does the risk estimate, which is why process changes need a risk review. In the root cause analysis guide, the same weak-lamination defect came from a new laminator running below its validated pressure.
From PFMEA to control plan
The control plan lists what is checked at each step, how, how often, and what happens when a check fails. Every control the PFMEA relies on should appear in it, and every line in it should trace back to a PFMEA row. If a control isn't in the control plan, nobody is required to do it.
| Step | Characteristic | Specification | Method | Frequency | Reaction if out of spec |
|---|---|---|---|---|---|
| Lamination | Nip roller pressure | Validated range on the process sheet | Inline pressure sensor with alarm | Continuous | Line stops; affected material segregated and assessed through the NC process |
| Lamination | Peel strength | Minimum force per the design spec | Peel test (validated method) | 5 samples per roll | Hold the lot; open an NC; check the process against the PFMEA |
| PCB and battery | Battery polarity and seating | Correct orientation, fully seated | Vision system | 100% | Reject and log; investigate if reject rate exceeds the alert limit |
| Housing weld | Seal integrity | No leak at test pressure | Leak test | 100% | Reject; trend weld parameters |
| Final test | Battery current draw in standby | Within the specified range | Calibrated current meter at the test station | 100% | Hold the lot; escalate to engineering |
Part 5 of 11 · Use-related risk · 6 min read
Use-Related Risk Analysis and Critical Tasks
A device that works perfectly can still hurt someone if it's used wrongly. A use-related risk analysis (URRA) looks at every task a user performs and asks what goes wrong when the user gets it wrong.
Use errors are a design problem
IEC 62366-1 defines a use error as a user action, or failure to act, that leads to a different result than the manufacturer intended or the user expected. It's distinct from abnormal use, a deliberate act that goes against normal use and is beyond reasonable design control.[10] Use errors are foreseeable, so they belong in the risk analysis like any other hazard source.
FDA's human factors guidance defines a critical task as one that, “if performed incorrectly or not performed at all, would or could cause serious harm”; FDA's human factors guidance counts compromised medical care as harm.[11] Critical tasks are what summative (validation) testing has to cover. In May 2026, FDA finalized its guidance on the human factors content of marketing submissions, which sorts submissions into three categories through a series of decision points about the user interface and its critical tasks.[12]
Build the URRA
Here are eight Orvelo tasks, from opening the pouch at home to a clinician reading the report. Severity drives whether a task is critical. Then choose a risk control for each: notice that only a design change that removes the task altogether takes it off the critical list. Everything else still has to be shown to work in validation testing. In this example, Serious and above count as serious harm; map your own severity scale to FDA's definition in your plan.
Try it
Orvelo use-related risk analysis
| Task | Possible use error → harm | Severity | Critical? | Risk control |
|---|
FDA's example URRA table also records the hazardous situation for each use error and how the effectiveness of each control will be validated.[12] The analysis is a living document: formative testing usually turns up use errors nobody predicted, and each one goes back into the table.
Where the URRA links in
A use error is one more cause on the way to a hazardous situation, so URRA rows link to the hazard analysis, to the same hazardous situations and harms the DFMEA links to, with severity from the same harms table. FDA's example URRA table has columns for the hazardous situation and the harm for this reason.[12] Most teams still keep the URRA as its own document, because it has to stand on its own in the human factors report.
The link to the DFMEA is indirect, and it runs through risk controls. When a use error is fixed by changing the design, the new design feature can fail too, and that failure belongs in the DFMEA. ISO 14971 calls this checking for risks introduced by risk controls.[1] Switch between the two designs to see it.
Try it
Start-button task: before and after the design change
Which human factors submission category?
FDA's 2026 guidance asks a short series of questions to decide what human factors information a submission needs. This simplified version follows its logic; check the guidance itself for the full decision points and examples.[12]
Try it
HF submission category helper
Answer the questions to see the category.
Part 6 of 11 · Software risk · 5 min read
Software Risk and IEC 62304 Safety Classes
Software doesn't wear out or fail at random, so the usual probability estimates don't apply. Instead, IEC 62304 sorts software by how bad the outcome of a failure could be, and sets the development rigor to match.
Why software is scored differently
A software defect is a design flaw: given the same inputs, it fails every time. There's no meaningful failure rate to put on the probability axis. IEC 62304, the software lifecycle standard, deals with this by assuming the probability of a software failure is 1 when assigning a software safety class.[13] The class depends on the worst harm that could result after considering risk controls outside the software system, such as an independent hardware cutoff:
Class A
The software can't contribute to a hazardous situation, or the risk is acceptable once controls outside the software are considered.
No unacceptable riskClass B
The software can contribute to a hazardous situation and the resulting risk isn't acceptable, but the worst possible harm is a non-serious injury.
Non-serious injuryClass C
As for B, but the worst possible harm is death or serious injury. Class C requires the most development and documentation rigor.
Death or serious injuryIEC 62304:2006 with Amendment 1:2015 is still the current edition. A second edition is in development, so check its status before relying on any draft text.
Classify the Orvelo software
Pick a software item and answer the questions. Watch what happens to the patch firmware when you add an independent hardware thermal cutoff: a control outside the software can lower the class, which can save a great deal of development effort.
Try it
Software safety class finder
FDA documentation level: a different question
For US submissions, FDA's 2023 guidance on device software functions asks for Basic or Enhanced documentation. Enhanced applies where a software failure “could present a hazardous situation with a probable risk of death or serious injury”, and the risk is assessed before any risk controls are applied. It also recommends Enhanced for some device types, including Class III devices.[14]
Same software, different answers
Because IEC 62304 allows external risk controls and FDA's documentation level doesn't, the Orvelo firmware with a hardware cutoff can be Class A under IEC 62304 and still need Enhanced documentation for FDA. Work out both, and record the rationale for each.
The class decides how much lifecycle process you owe: architecture, detailed design, unit verification. Changing it late means redoing work.
Libraries and operating systems you didn't write (SOUP, in IEC 62304 terms) can contribute to hazardous situations too. List their known anomalies and how you'll monitor them; Part 7 covers the security side.
Software causes sit on the same chains as hardware causes. "Firmware fails to shut down the circuit during a battery fault" belongs on the overheating chain in the hazard analysis, not in a separate software-only document.
Part 7 of 11 · Cybersecurity · 7 min read
Cybersecurity Risk: Threats, Exploitability and Patient Safety
A connected device has a second kind of hazard source: people who want to misuse it. Security risk is assessed in its own process, then any threat that could hurt a patient is carried into the safety risk analysis.
Why security risk needs its own analysis
Random failures have rates you can estimate. Attacks don't: whether someone attacks your device depends on motive, opportunity and how easy the attack is, and that changes as new vulnerabilities are published. FDA's premarket cybersecurity guidance therefore treats security risk management as a separate process from safety risk management, and recommends assessing exploitability (how easily a vulnerability could be used) instead of probability.[15] AAMI TIR57 and AAMI SW96 describe how to run that process alongside ISO 14971.[16][17]
The two processes meet at the point of harm. If a threat could cause a hazardous situation, such as a patch that silently stops recording, it goes into the hazard analysis and is scored there like any other cause. Threats that only affect privacy or business operations stay in the security risk assessment, where they still need controls.
What regulators ask for
US: Section 524B of the FD&C Act applies to “cyber devices”: devices that include software, can connect to the internet and have characteristics that could be vulnerable to cybersecurity threats. Submissions must include a plan to monitor, identify and address postmarket vulnerabilities, including coordinated vulnerability disclosure; processes and procedures that “provide a reasonable assurance that the device and related systems are cybersecure”, including making updates and patches available on a regular cycle and sooner for critical vulnerabilities; and a software bill of materials (SBOM).[18][15]
EU: MDR Annex I covers information security in sections 17.2 and 17.4, and protection against unauthorized access in section 18.8. MDCG 2019-16 is the EU guidance on how to meet them.[6][19] IEC 81001-5-1 sets out security activities across the software lifecycle and is recognized by FDA.[20]
Score the Orvelo threats
Five threat scenarios for the patch, its app and its cloud service. Each is placed by exploitability and impact. Turn on security controls and watch the scenarios move. The scenarios marked with a heart could affect a patient, so they also go into the hazard analysis.
Try it
Security risk assessment
Security controls
Exploitability × impact
| ID | Threat scenario | Exploitability | Security risk | Hazard analysis? |
|---|
The scales and zones here are illustrative. Define your own security risk acceptance criteria in your security risk management plan, separately from the safety criteria in Part 1.
A CVSS score is not a risk score
When a vulnerability is published in a component you use, it usually comes with a CVSS score. CVSS describes how severe the vulnerability is in general; it doesn't know how your device uses the component or what a patient would experience.[21] FDA has recognized CVSS v4.0 as a consensus standard, and a rubric developed by MITRE adapts the earlier CVSS v3.0 for medical devices.[22][23] Either way, the score is an input to your risk assessment, not the answer.
Try it
Triage a published vulnerability
FDA's postmarket cybersecurity guidance asks manufacturers to decide whether the risk to essential clinical performance is controlled (acceptable residual risk) or uncontrolled. The answer decides how quickly you must act and whether customers and FDA need to be told.[24]
Making it work in practice
Draw the system: patch, phone, app, cloud, clinician portal, update server. Mark every interface and trust boundary, then ask what an attacker could do at each one. Threats come from the model, not from a generic checklist.
The SBOM is how you find out a new vulnerability affects you. Match it against vulnerability feeds continuously, not only at submission.
Give each safety-relevant threat an ID that appears in both the security risk assessment and the hazard analysis. When a security control changes, the safety team should see it.
Verify security controls with penetration testing and fuzzing of the interfaces in the threat model, as well as by reviewing the design.
Part 8 of 11 · AI/ML risk · 7 min read
AI and Machine Learning Risk: Thresholds, Bias and Drift
A machine learning model is never simply right or wrong. It makes a trade-off between missed cases and false alarms, it can perform worse for some patients than others, and its performance can change after release. Each of these is a risk to analyze.
What changes with machine learning
ISO 14971 still applies to AI-enabled devices. In June 2026, ISO published ISO/TS 24971-2, guidance on applying ISO 14971 to machine learning. It builds on AAMI TIR34971 and covers hazards that come from data and from the way people use model outputs. It excludes large language models and generative AI.[25][26] Typical sources of harm include:
- Performance trade-offs: the decision threshold sets how many true cases are missed and how many false alarms are raised.
- Data problems: training data that doesn't represent the people or conditions the device will meet in use.
- Automation bias: users trusting the output too much, or ignoring it after too many false alarms.
- Change over time: performance drifts when patients, sensors or software around the model change.
Regulatory landscape (October 2026)
US: FDA finalized its guidance on predetermined change control plans (PCCPs) for AI-enabled device software functions in December 2024; the current version is dated August 2025. A PCCP lets a manufacturer describe planned model changes, and how they'll be validated, in the original submission.[27] FDA's broader guidance on AI-enabled device software functions was issued as a draft in January 2025.[28] The good machine learning practice principles from FDA, Health Canada and the MHRA have been developed further by IMDRF.[29][30]
EU: AI systems that are medical devices, or safety components of them, needing notified body assessment are high-risk under the AI Act. An amending regulation adopted in 2026 moved the date their high-risk obligations apply to 2 August 2028.[31][32]
Choose the AF detection threshold
The Orvelo cloud model gives each recording a score from 0 to 1 for how likely it is to show atrial fibrillation (AF). Recordings above the threshold are flagged for a clinician. Move the threshold and watch the two kinds of error trade off. Then switch to the subgroup with noisy, low-amplitude signals, which were scarce in the training data.
Try it
Threshold and subgroup explorer
Each kind of error leads to a different harm. A missed AF episode can delay anticoagulation and leave stroke risk untreated; a false alarm costs clinician time and can cause anxiety or unnecessary tests. The risk analysis should state which error matters more, set an acceptance criterion for each, and check it in every relevant subgroup, not just overall.
The score distributions are simulated for illustration and don't describe any real algorithm. Each curve shows the shape of the scores for its group; the counts assume 10% of recordings contain AF.
Watch for drift after release
In this example, a phone operating system update in month 10 changes how the app compresses ECG data before upload. The model was never trained on the new signal, and its sensitivity starts to slide. How quickly would you notice? Choose how often you compare model results with clinician-confirmed diagnoses.
Try it
Post-market performance monitoring
Decide in the risk management plan what minimum sensitivity and specificity are acceptable, and for which subgroups, before you see the test results.
Data used to evaluate the model must be kept separate from training data, or the results will look better than real-world performance.
Phone operating systems, sensors and patient populations all change. List them in the risk analysis as causes of degraded performance, and monitor for each.
If you expect to retrain, describe the change and its validation in a PCCP, so a fix for drift doesn't require a new submission each time.
Part 9 of 11 · Risk control · 12 min read
Risk Control: Design, Protective Measures and Information for Safety
Analysis tells you which risks need reducing. Risk control is where you reduce them: choosing controls in the right order, recording why, proving each one works, and checking that none of them creates a new problem.
Three kinds of control, in order of preference
ISO 14971 asks manufacturers to consider risk control options in a fixed order, and to use them in that order where they're practicable.[1] The EU MDR sets out the same order in Annex I, section 4.[6] The order reflects how much each kind of control depends on people doing the right thing.
1. Inherent safety by design and manufacture
Remove the hazard, or make the hazardous situation less likely or less severe at its source. Works without anyone noticing or acting.
Orvelo: a coin cell that can't heat the surface to burn temperature even when shorted; auto-start on skin contact.
2. Protective measures
Detect a problem and respond to it, in the device itself or in the manufacturing process. The hazard is still there; the protection stands between it and the patient.
Orvelo: a thermal cutoff; a “check patch” alert; a 100% leak test on the production line.
3. Information for safety
Warnings, precautions, contraindications and instructions, plus training where appropriate. It only works if the user reads it, understands it, remembers it and acts on it.
Orvelo: “Remove the patch if it feels warm”; “Do not apply over broken skin.”
The order is not “use only one kind.” Most serious hazards end up with controls from more than one category, as the bow-tie at the end of this part shows. What the order rules out is reaching for a warning label when a design change was practicable.
Which kind of control is it?
The boundaries trip people up, especially for controls in manufacturing and for training. Sort these nine Orvelo controls.
Try it
Sort the controls
Run a control option analysis
For each hazardous situation, the file should show which options you considered, which you chose and why. Pick a hazardous situation, choose controls, and watch the residual risk, the verification you now owe, and any new risks the controls bring with them. The probabilities and their reductions are illustrative.
Try it
Control option analysis
Verify implementation and effectiveness
ISO 14971 asks for two different things for every control: evidence that it was implemented, and evidence that it is effective in reducing the risk.[1] Teams often have the first and assume the second.
| Control | Implemented: is it really in the device? | Effective: does it reduce the risk? |
|---|---|---|
| Thermal cutoff | Part on the released BOM and schematic; requirement traced to the design output | Fault test: with the battery shorted, the surface stays below the burn limit across samples and temperatures |
| “Check patch” alert | Software requirement implemented and verified | The alert fires reliably when skin contact is lost, and users respond to it in human factors validation |
| 100% leak test | Test step in the released manufacturing procedure; validated test method | Leak-test escapes tracked; field water-ingress complaints stay within the estimate |
| IFU warning | Approved text in the released IFU and on the pouch | Users understand and act on it in human factors validation, not just that the text exists |
Effectiveness evidence for information for safety usually comes from human factors work: knowledge-task questions and observed behavior in summative testing, which is why the URRA in Part 5 and this step are so closely linked.
Check the controls themselves
ISO 14971 asks whether each control introduces new hazards or hazardous situations, or changes risks already estimated.[1] A thermal cutoff that trips falsely stops recording. A lower-energy cell may run out before the end of the wear period. Each becomes a new row in the analysis.
Before moving on, check that every identified hazardous situation has been dealt with and that no risk control activity is still open.[1] A trace report that lists any hazardous situation without an evaluated residual risk, and any control without implementation and effectiveness evidence, does this in seconds.
Design verification proves a control works; it doesn't reduce risk by itself. “Verified by test report TR-112” belongs in the verification column, not the control column. A production test that screens every unit is different: it's a protective measure in manufacturing.
Telling users that a risk exists (“mild skin redness is common”) is disclosure of residual risk, covered in Part 10. Information for safety tells users what to do or avoid.
See every barrier at once: the bow-tie
An FMEA spreads one hazard across many rows. A bow-tie pulls it back together, which makes it a good final check on the controls for your most severe hazards.
IEC 31010, the general standard on risk assessment techniques, describes the bow-tie as a diagram centered on an unwanted event. Threats that could cause the event sit on the left, with preventive barriers between them and the event. Consequences sit on the right, with mitigative barriers that limit how bad the outcome gets.[33] The left side reads like a simplified fault tree and the right side like an event tree.
A bow-tie doesn't produce a risk score. Its strength is communication: in a design review, anyone can see which paths have two independent barriers and which have one, or none.
The central event is the Orvelo circuit producing excess heat while the patch is worn. Click any barrier to remove it or put it back. Then make the thermal cutoff firmware-controlled and see what happens to the firmware path.
Try it
Orvelo overheating bow-tie
| Path | Working barriers | Status |
|---|
Getting value from a bow-tie
Two barriers that share a cause of failure count as one. The firmware-controlled cutoff above fails exactly when the firmware does.
Conditions that weaken a barrier belong on the diagram. Here, a sleeping patient defeats “remove the patch if it feels warm”.
A bow-tie for every hazard is too much. Draw them for the few events with the worst consequences, and use them to explain the risk controls to reviewers and auditors.
Each barrier should trace to a risk control in the hazard analysis or FMEA, with its verification. The diagram summarizes the file; it doesn't replace it.
Part 10 of 11 · Residual risk and benefit-risk · 9 min read
Residual Risk, Benefit-Risk and the Risk Management Review
Every control leaves some risk behind. The last stage of ISO 14971 decides whether what's left is acceptable, one hazardous situation at a time and then for the device as a whole, and weighs it against what the device does for patients.
Residual risk, one hazardous situation at a time
After controls are in place, re-estimate each risk with the same method you used before, and evaluate it against the criteria in your plan. If it's not acceptable, apply more control. Only when further control isn't practicable does ISO 14971 let you turn to a benefit-risk analysis for that risk; if the benefits don't outweigh it, it stays unacceptable.[1] Walk a residual risk through the decision.
Try it
What happens to a residual risk?
When does reducing a risk stop? That depends on which principle your policy follows, and the answer has changed over the past 25 years. The next section explains.
ALARP, AFAP and how the rules evolved
Three phrases describe how far risk reduction has to go. They sound alike and lead to different files.
| Principle | Where it comes from | When can you stop reducing? | Can cost count? |
|---|---|---|---|
| ALARP As low as reasonably practicable | UK health and safety law. In Edwards v National Coal Board (1949), the court said the risk must be weighed against the sacrifice in money, time or trouble needed to avert it.[34] | When the cost of further reduction would be grossly disproportionate to the risk reduction gained. The presumption is to reduce. | Yes |
| ALARA As low as reasonably achievable | Radiation protection, where it describes keeping doses as low as reasonably achievable. | In practice, much like ALARP. | Yes |
| AFAP As far as possible | The EU medical device directives, carried into the MDR, which defines it as reducing risks “as far as possible without adversely affecting the benefit-risk ratio”.[6] | When no further control is possible given the state of the art, or further control would worsen the benefit-risk balance. | No |
The “No” for AFAP is the prevailing EU interpretation, which comes from the harmonized standard's annexes and notified body practice; the MDR itself doesn't mention cost.[35] Many risk matrices have three zones because of ALARP: broadly acceptable, a middle region where risk is reduced “as low as reasonably practicable”, and unacceptable. Under AFAP, the middle region is no longer a place where you can stop on cost grounds, and even the green zone needs evidence that the risk was reduced as far as possible.
Explore
How the rules evolved
The seven deviations, then and now
The 2012 European edition of ISO 14971 kept the 2007 text but added annexes (Annexes Z) listing seven points where the standard fell short of the directives. Notified bodies and industry associations published a consensus paper in 2014 on how to apply them.[35] The headings below are shortened from those annexes. Most of the friction has since been resolved by the MDR's wording and the 2019 edition, but the ideas behind the deviations still shape what EU reviewers expect.
| Deviation (EN ISO 14971:2012) | What it meant | Where it stands now |
|---|---|---|
| 1. Treatment of negligible risks | All risks must be reduced, not just those above a negligible threshold. | Still the EU expectation: AFAP applies to every risk. |
| 2. Manufacturer discretion on acceptability | Acceptability criteria can't be set freely; they must meet the legal requirements. | The EN 2021 amendment says the policy must ensure the criteria comply with the MDR's safety and performance requirements.[36] |
| 3. AFAP versus ALARP | Risks must be reduced as far as possible, without economic considerations. | The MDR defines AFAP. ISO 14971:2019 describes ALARP-, ALARA- and AFAP-style approaches the policy can choose, so EU products follow AFAP.[1] |
| 4. Discretion on benefit-risk analysis | A benefit-risk analysis is always needed, not only when a risk is unacceptable. | ISO 14971:2019 always weighs the overall residual risk against the benefits (Clause 8).[1] |
| 5. Discretion on risk control options | All practicable control options must be used, not just enough to reach “acceptable”. | MDR Annex I, section 4 sets the order; the AFAP expectation means considering every option. |
| 6. The first control option | The first option must be inherently safe design and manufacture, as the directives required. | Aligned: ISO 14971:2019 lists inherent safety by design and manufacture first. |
| 7. Information to users and residual risk | Information given to users couldn't be counted as reducing risk further. | The MDR lists information for safety as the third control option, and requires users to be told about residual risks.[6] |
Try it
Same residual risks, two principles
| ID | Residual risk | Zone | What the file needs |
|---|
Many companies apply AFAP everywhere, so one risk management file serves every market. Whatever you choose, state it in the policy and apply it consistently.
Benefit-risk analysis
Benefit-risk shows up twice in ISO 14971: for an individual residual risk that can't be reduced further, and for the overall residual risk of the device, which is always weighed against the benefits of the intended use.[1][4] A benefit has to be a positive impact on health: a better diagnosis, a better outcome, better patient management, backed by clinical data. A feature or a lower price isn't a medical benefit.
FDA's benefit-risk guidance for PMA and De Novo decisions lists the factors reviewers weigh, and it's a useful checklist for any device:[37]
Benefits
The type, magnitude, probability and duration of the benefit.
Risks
The severity, types, number and rates of harmful events, their probability and duration, and for diagnostics, the risk of false positives and false negatives.
Other factors
Uncertainty, patient-centric assessments and patient-reported outcomes, the disease, patient perspectives, alternative treatments or diagnostics, risk mitigation, postmarket data, and unmet medical need.
Try it for the Orvelo, compared with the alternative it would replace: a 24-hour Holter monitor. All figures are illustrative and per 1,000 patients monitored.
Try it
Benefit-risk worksheet: Orvelo vs a 24-hour Holter
The worksheet structures the comparison; it doesn't make the decision. The conclusion is a documented judgment by people with clinical knowledge, based on the clinical evaluation.
Overall residual risk
Individually acceptable risks can add up. Clause 8 of ISO 14971 asks you to judge the device's overall residual risk against the benefits of its intended use, considering how all the residual risks combine, with a method and criteria fixed in advance in your plan.[1] ISO/TR 24971 offers several possible approaches, including reviewing data on similar devices already on the market and judgment by a cross-functional team with relevant clinical knowledge, and notes that the criteria for overall residual risk can differ from those for individual risks.[4]
Try it
Orvelo overall residual risk
Disclose significant residual risks
ISO 14971 asks you to inform users of significant residual risks in the accompanying documentation, and the MDR requires manufacturers to inform users of any residual risks.[1][6] Disclosure is different from information for safety: it describes what can still happen, so users and patients can make informed decisions. Information for safety tells them what to do.[4]
| Disclosure of residual risk (describes) | Information for safety (instructs) |
|---|---|
| Mild redness or itching under the adhesive is common and usually fades within a few days of removal. | Do not apply the patch over broken or irritated skin. |
| Short periods without ECG data can occur if the patch loses contact with the skin. | Check data coverage in the report before concluding that no arrhythmia occurred. |
| The algorithm can miss some episodes of atrial fibrillation. | Review the full ECG for patients with symptoms, whatever the algorithm reports. |
The risk management review and report
Before commercial release, ISO 14971 requires a review of how the risk management plan was carried out. At a minimum, the review confirms three things, and its results are recorded as the risk management report:[1]
The plan was implemented
Every activity in the risk management plan was done, by the people and at the points it named, and the records exist.
The overall residual risk is acceptable
The clause 8 evaluation is complete, with its benefit-risk conclusion and the significant residual risks identified for disclosure.
Production and post-production methods are in place
There's a working way to collect and review information from production, complaints, service and the literature, and to feed it back into the analysis. Part 11 shows that loop in action.
The report isn't the end of the file. It's a snapshot that's updated whenever the risk picture changes: a design change, a new hazard from the field, or a change in the state of the art.
Part 11 of 11 · Bringing it together · 6 min read
Bringing It Together: One Risk Management File
Each tool in this guide answers one question. The risk management file is where those answers connect, and post-market data is how you find out whether they were right.
Which tool answers which question
| Tool | The question it answers | Main reference |
|---|---|---|
| Hazard analysis | How could this device harm someone, and how likely and severe is that harm? | ISO 14971 |
| Design FMEA | How could each part of the design fail, and what happens if it does? | IEC 60812 |
| Process FMEA | How could manufacturing produce a defective device, and would we catch it? | IEC 60812 |
| URRA | What happens when a user makes a mistake, and which tasks are critical? | IEC 62366-1 |
| Software classification | How much development rigor does each software item need? | IEC 62304 |
| Security risk assessment | How could an attacker misuse the device, and could that harm a patient? | AAMI SW96, TIR57 |
| ML risk analysis | How could the model’s errors, data or drift lead to harm? | ISO/TS 24971-2 |
| Risk control option analysis | Which controls should we use, in what order, and do they work? | ISO 14971 |
| Bow-tie | Does every path to a serious harm have enough independent barriers? | IEC 31010 |
| Benefit-risk analysis | Do the device’s benefits outweigh the risk that remains? | ISO 14971, ISO/TR 24971 |
Is the field telling you something?
The hazard analysis estimated how often patients would be burned. Complaints show how often it actually happens, at least the cases people report. ISO 14971 requires you to collect and review this information and act if a risk estimate no longer holds.[1] Enter a year of field data and compare.
Try it
Field rate against the estimate
When the field rate is higher than the estimate, the reason matters as much as the number. The Pareto analysis and fault tree analysis chapters of our root cause analysis guide show how to find it, and the findings go back into the hazard analysis and FMEAs.
Risk management file checklist
A quick self-check against the main clauses of ISO 14971:2019. Tick what your file already has; the gaps list shows what's left.[1]
Try it
Is the file complete?
The thread that runs through it all
Every tool in this guide ends the same way: a hazardous situation, an estimate of its risk, a control, and evidence that the control works. Keep those links intact, from the first hazard to the latest complaint, and the risk management file becomes a working tool for design decisions rather than a document assembled before a submission.
Next: when the field talks back
Six months after launch, three things land in the same week: a cluster of complaints about warm patches from one lot, a published case report on adhesive dermatitis with a competitor's patch, and a phone update that quietly changes how ECG data is compressed.
Each one could change a risk estimate. But which belong in the risk management file, which stay in the post-market surveillance system, and who decides whether the burn estimate from Part 2 still holds? The line is clearer than most files make it look, and getting it wrong is how the risk file and the PMS reports end up telling different stories.
Our post-market surveillance guide follows the Orvelo into the field: complaint trending, vigilance, PSURs, post-market clinical follow-up, and how each one flows back into the risk file. Read it now →
References
- ISO 14971:2019, Medical devices — Application of risk management to medical devices. International Organization for Standardization. www.iso.org/standard/72704.html
- U.S. FDA, Inspection Observations: Form 483 observation spreadsheets by fiscal year, Devices tab, FY2021–FY2025. www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/inspection-references/inspection-observations
- ISO 10993-1:2025, Biological evaluation of medical devices — Part 1: Requirements and general principles for the evaluation of biological safety within a risk management process, 6th edition. International Organization for Standardization, November 2025. www.iso.org/standard/84512.html
- ISO/TR 24971:2020, Medical devices — Guidance on the application of ISO 14971 (Annex B, risk analysis techniques). www.iso.org/standard/74437.html
- 21 CFR Part 820, Quality Management System Regulation. Electronic Code of Federal Regulations. www.ecfr.gov/current/title-21/chapter-I/subchapter-H/part-820
- Regulation (EU) 2017/745 on medical devices (MDR). EUR-Lex. eur-lex.europa.eu/eli/reg/2017/745/oj
- Regulation (EU) 2017/746 on in vitro diagnostic medical devices (IVDR). EUR-Lex. eur-lex.europa.eu/eli/reg/2017/746/oj
- IEC 60812:2018, Failure modes and effects analysis (FMEA and FMECA), Edition 3.0. International Electrotechnical Commission. webstore.iec.ch/en/publication/26359
- AIAG & VDA, FMEA Handbook, 1st edition. Automotive Industry Action Group and Verband der Automobilindustrie, 2019. webshop.vda.de/QMC/en/aiag-vda-fmea-handbook_eng
- IEC 62366-1:2015+AMD1:2020, Medical devices — Part 1: Application of usability engineering to medical devices (consolidated version 1.1). International Electrotechnical Commission. webstore.iec.ch/en/publication/67220
- U.S. FDA, Applying Human Factors and Usability Engineering to Medical Devices. Final guidance, February 2016; revised August 2026. www.fda.gov/regulatory-information/search-fda-guidance-documents/applying-human-factors-and-usability-engineering-medical-devices
- U.S. FDA, Content of Human Factors Information in Medical Device Marketing Submissions. Final guidance, May 2026. www.fda.gov/regulatory-information/search-fda-guidance-documents/content-human-factors-information-medical-device-marketing-submissions
- IEC 62304:2006+AMD1:2015, Medical device software — Software life cycle processes. International Electrotechnical Commission. webstore.iec.ch/en/publication/22794
- U.S. FDA, Content of Premarket Submissions for Device Software Functions. Final guidance, June 2023. www.fda.gov/regulatory-information/search-fda-guidance-documents/content-premarket-submissions-device-software-functions
- U.S. FDA, Cybersecurity in Medical Devices: Quality Management System Considerations and Content of Premarket Submissions. Final guidance, February 2026. www.fda.gov/regulatory-information/search-fda-guidance-documents/cybersecurity-medical-devices-quality-management-system-considerations-and-content-premarket
- AAMI TIR57:2016/(R)2023, Principles for medical device security — Risk management. Association for the Advancement of Medical Instrumentation. array.aami.org/doi/10.2345/9781570206122
- ANSI/AAMI SW96:2023, Standard for medical device security — Security risk management for device manufacturers. Association for the Advancement of Medical Instrumentation. array.aami.org/doi/10.2345/9781570208621
- Federal Food, Drug, and Cosmetic Act § 524B, Ensuring cybersecurity of devices, 21 U.S.C. § 360n-2. www.law.cornell.edu/uscode/text/21/360n-2
- Medical Device Coordination Group, MDCG 2019-16 Rev.1: Guidance on cybersecurity for medical devices. December 2019; Rev.1 July 2020. health.ec.europa.eu/document/download/b23b362f-8a56-434c-922a-5b3ca4d0a7a1_en
- IEC 81001-5-1:2021, Health software and health IT systems safety, effectiveness and security — Part 5-1: Security — Activities in the product life cycle. International Electrotechnical Commission. webstore.iec.ch/en/publication/63293
- FIRST, Common Vulnerability Scoring System version 4.0: Specification Document. Forum of Incident Response and Security Teams, November 2023. www.first.org/cvss/specification-document
- U.S. FDA, Recognized Consensus Standards database: Common Vulnerability Scoring System version 4.0, recognition no. 13-140. www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfStandards/detail.cfm?standard__identification_no=47094
- MITRE, Rubric for Applying CVSS to Medical Devices. Qualified by FDA as a Medical Device Development Tool, 2020. www.mitre.org/news-insights/publication/rubric-applying-cvss-medical-devices
- U.S. FDA, Postmarket Management of Cybersecurity in Medical Devices. Final guidance, December 2016. www.fda.gov/regulatory-information/search-fda-guidance-documents/postmarket-management-cybersecurity-medical-devices
- ISO/TS 24971-2:2026, Medical devices — Guidance on the application of ISO 14971 — Part 2: Machine learning in artificial intelligence. International Organization for Standardization, June 2026. www.iso.org/standard/24971-2
- AAMI TIR34971:2023, Application of ISO 14971 to machine learning in artificial intelligence — Guide. Association for the Advancement of Medical Instrumentation. array.aami.org/doi/book/10.2345/9781570208669
- U.S. FDA, Marketing Submission Recommendations for a Predetermined Change Control Plan for Artificial Intelligence-Enabled Device Software Functions. Final guidance, December 2024; current version August 2025. www.fda.gov/regulatory-information/search-fda-guidance-documents/marketing-submission-recommendations-predetermined-change-control-plan-artificial
- U.S. FDA, Artificial Intelligence-Enabled Device Software Functions: Lifecycle Management and Marketing Submission Recommendations. Draft guidance, January 2025. www.fda.gov/regulatory-information/search-fda-guidance-documents/artificial-intelligence-enabled-device-software-functions-lifecycle-management-and-marketing
- U.S. FDA, Health Canada and UK MHRA, Good Machine Learning Practice for Medical Device Development: Guiding Principles. October 2021. www.fda.gov/media/153486/download
- IMDRF, Good machine learning practice for medical device development: Guiding principles (IMDRF/AIML WG/N88 FINAL:2025). International Medical Device Regulators Forum, January 2025. www.imdrf.org/sites/default/files/2025-02/IMDRF_AIML%20WG_GMLP_N88%20Final.pdf
- Regulation (EU) 2024/1689 laying down harmonised rules on artificial intelligence (Artificial Intelligence Act). EUR-Lex. eur-lex.europa.eu/eli/reg/2024/1689/oj
- Regulation (EU) 2026/1744 amending Regulation (EU) 2024/1689 and other acts (Digital Omnibus on AI). Official Journal, 24 July 2026. EUR-Lex. eur-lex.europa.eu/eli/reg/2026/1744/oj
- IEC 31010:2019, Risk management — Risk assessment techniques, Edition 2.0. International Electrotechnical Commission. webstore.iec.ch/en/publication/59809
- UK Health and Safety Executive, ALARP “at a glance”. www.hse.gov.uk/enforce/expert/alarpglance.htm
- Notified Bodies Recommendation Group, Consensus paper for the interpretation and application of Annexes Z in EN ISO 14971:2012, version 1.1, October 2014. www.team-nb.org/wp-content/uploads/2017/08/NBRG_WG-RM_Interim_NBmed_Consensus_Version_140812_1_1.pdf
- BSI, European standard for risk management of medical devices is amended (EN ISO 14971:2019+A11:2021). compliancenavigator.bsigroup.com/en/medicaldeviceblog/european-standard-for-risk-management-of-medical-devices-is-amended/
- U.S. FDA, Factors to Consider When Making Benefit-Risk Determinations in Medical Device Premarket Approval and De Novo Classifications. Final guidance, August 2019. www.fda.gov/regulatory-information/search-fda-guidance-documents/factors-consider-when-making-benefit-risk-determinations-medical-device-premarket-approval-and-de
The Orvelo wearable ECG patch is fictional. Its risk scores, probabilities, task lists, test data and other figures are illustrative and do not describe any real product or company. Severity and probability scales are examples; each manufacturer defines its own in its risk management plan.