EU Harmonises EN ISO 14155:2026 and EN ISO 11137-1:2026 Under the MDR and IVDR
Why it matters
On 10 November 2026, the Official Journal published two Commission Implementing Decisions dated 9 November 2026 [3]. They update the lists of harmonised standards under the Medical Devices Regulation (MDR, Regulation (EU) 2017/745) and the IVD Regulation (IVDR, Regulation (EU) 2017/746) [1][2]. Both Decisions entered into force on the day of publication [1][2].
The most important changes affect two standards: EN ISO 14155:2026 on good clinical practice for clinical investigations, and EN ISO 11137-1:2026 on radiation sterilization. Compliance with a harmonised standard gives a presumption of conformity with the corresponding requirements from the date its reference is published [2]. Technical documentation, clinical investigation plans and sterilization validation files that cite these standards will need review.
What changed
Status: NEW (in force from 10 November 2026).
- MDR list (Decision (EU) 2026/2479, amending Decision (EU) 2021/1182). EN ISO 11137-1:2026 replaces EN ISO 11137-1:2015 and its A2:2019 amendment. EN ISO 14155:2026 replaces EN ISO 14155:2020 and its A11:2024 amendment. Seven further references are added [1].
- IVDR list (Decision (EU) 2026/2481, amending Decision (EU) 2021/1195). EN ISO 11137-1:2026 replaces EN ISO 11137-1:2015 and its A2:2019 amendment. This is the only change to the IVDR list [2].
- Deferred withdrawal. The superseded EN ISO 11137-1:2015 and EN ISO 14155:2020 references are deleted with effect from 10 May 2028. This gives manufacturers time to adapt [1][2].
Recitals in both Decisions explain the process. The Commission asked CEN and Cenelec to revise the standards. It then assessed, together with them, whether the superseding standards meet the request and the relevant MDR or IVDR requirements before publishing the references [1][2]. Because the old references stay listed until 10 May 2028, manufacturers can claim presumption of conformity under either edition during the transition INTERPRETATION based on the deferred deletion. After that date, only the 2026 editions will appear on the lists.
The seven other references are added to the MDR list as new entries rather than as replacements: two parts of the ISO 10993 biological evaluation series, small-bore connectors, washer-disinfectors for non-critical devices, two medical electrical equipment particular standards and latex condoms [1]. They confer presumption of conformity under the MDR from 10 November 2026 [1][2].
What the two Decisions change
Annexes to Implementing Decisions (EU) 2026/2479 and 2026/2481, OJ L 10.11.2026 [1][2].
| Standard | Subject | Regulation | Change |
|---|---|---|---|
| EN ISO 11137-1:2026 | Radiation sterilization, Part 1 | MDR and IVDR | Supersedes 2015 + A2:2019; old reference withdrawn 10 May 2028 |
| EN ISO 14155:2026 | Clinical investigation, good clinical practice | MDR | Supersedes 2020 + A11:2024; old reference withdrawn 10 May 2028 |
| EN ISO 10993-6:2026 | Biological evaluation: local effects after implantation | MDR | New reference |
| EN ISO 10993-7:2026 | Biological evaluation: ethylene oxide sterilization residuals | MDR | New reference |
| EN ISO 80369-1:2026 | Small-bore connectors: general requirements | MDR | New reference |
| EN ISO 15883-6:2026 | Washer-disinfectors: thermal, non-critical devices | MDR | New reference |
| EN IEC 60601-2-22:2020 + A11:2026 | Medical electrical equipment: laser equipment | MDR | New reference |
| EN IEC 60601-2-57:2026 + A11:2026 | Medical electrical equipment: non-laser light sources | MDR | New reference |
| EN ISO 4074:2026 | Natural rubber latex male condoms | MDR | New reference |
MDR list = Implementing Decision (EU) 2021/1182. IVDR list = Implementing Decision (EU) 2021/1195. Only EN ISO 11137-1:2026 is added to the IVDR list.
Who is affected
Medical devices (MDR). Sponsors and manufacturers running clinical investigations, because EN ISO 14155:2026 is now the harmonised GCP standard [1]. Manufacturers of radiation-sterilized devices (EN ISO 11137-1:2026). Manufacturers of implants and ethylene-oxide-sterilized devices (EN ISO 10993-6/-7:2026). Manufacturers of devices with small-bore connectors, washer-disinfectors, and laser or light-source equipment [1].
IVDs (IVDR). Only manufacturers of radiation-sterilized IVDs or accessories, such as sterile collection devices, are affected through EN ISO 11137-1:2026 [2]. INTERPRETATION EN ISO 14155:2026 was added only to the MDR list. Decision 2026/2481 does not change the harmonised status of any standard used for IVD performance studies.
Key dates
| Date | Event | Source status |
|---|---|---|
| 9 Nov 2026 | Commission adopts Decisions (EU) 2026/2479 and 2026/2481 | VERIFIED [1][2] |
| 10 Nov 2026 | Published in the OJ L series; both Decisions enter into force | VERIFIED [1][2][3] |
| 10 May 2028 | Deletion of the superseded EN ISO 11137-1:2015 and EN ISO 14155:2020 references applies | VERIFIED [1][2] |
Key dates
Dates from Implementing Decisions (EU) 2026/2479 and 2026/2481 [1][2].
- 9 Nov 2026Commission adopts both Implementing Decisions
- 10 Nov 2026Published in OJ L and enter into force; new references confer presumption of conformity
- 10 May 2028Deletion of superseded EN ISO 11137-1:2015 and EN ISO 14155:2020 references applies
Between 10 Nov 2026 and 10 May 2028 both the superseded and the superseding references are listed (INTERPRETATION based on the deferred deletion).
What RA teams should do
- Update standards lists. Record the new references, with the 10 May 2028 withdrawal date, in your harmonised-standards register and General Safety and Performance Requirements (GSPR) checklists.
- Gap-assess EN ISO 14155:2026. Compare clinical investigation SOPs, plan templates and monitoring practices with the 2026 edition before May 2028. Ongoing investigations need a documented transition rationale INTERPRETATION.
- Review sterilization files. Radiation sterilization validations and supplier agreements that cite EN ISO 11137-1:2015 should be assessed against the 2026 edition.
- Check biocompatibility. Implant and ethylene oxide residual evaluations should be compared with EN ISO 10993-6:2026 and 10993-7:2026 where these standards are relied on.
- Refresh supplier controls. Ask contract sterilizers and test laboratories when they will move to the 2026 editions, and update quality agreements and audit plans to match.
- Talk to your notified body. Agree how and when updated standards will be reflected in technical documentation INTERPRETATION.
References
- European Commission. Commission Implementing Decision (EU) 2026/2479 of 9 November 2026 amending Implementing Decision (EU) 2021/1182 as regards harmonised standards for sterilization of health care products, clinical investigation of medical devices for human subjects, natural rubber latex male condoms, biological evaluation of medical devices, washer-disinfectors, medical electrical equipment and small-bore connectors for liquids and gases in healthcare applications. OJ L, 2026/2479, 10.11.2026. eur-lex.europa.eu/legal-content/EN/TXT/?uri=OJ:L_202602479
- European Commission. Commission Implementing Decision (EU) 2026/2481 of 9 November 2026 amending Implementing Decision (EU) 2021/1195 as regards the harmonised standard for sterilization of health care products. OJ L, 2026/2481, 10.11.2026. eur-lex.europa.eu/legal-content/EN/TXT/?uri=OJ:L_202602481
- Publications Office of the EU. Official Journal L series, daily view, 10 November 2026. eur-lex.europa.eu/oj/daily-view/L-series/default.html?ojDate=10112026
Disclaimer: This post is regulatory intelligence for information only and is not legal or regulatory advice. Harmonised standards are voluntary means of showing conformity. Always confirm the current harmonised-standards list in the Official Journal.