Explainer · Quality system · 16 min read
DHF, DMR and DHR Are Gone: What the QMSR Calls Them Now
On 2 February 2026, the FDA's Quality Management System Regulation replaced the old Quality System Regulation, and with it went the design history file, the device master record, the device history record and the quality system record. The content didn't disappear. It moved into the records ISO 13485 asks for. Here's where everything went, what actually changed, and what to do about your own files.
What happened, in plain terms
For almost 30 years, US device manufacturers followed the Quality System Regulation (21 CFR Part 820, in force since 1997). It had its own vocabulary, including four named compilations of records: the design history file, the device master record, the device history record and the quality system record.[1]
In February 2024, FDA published the final rule that rewrote Part 820 as the Quality Management System Regulation, or QMSR. It became effective on 2 February 2026.[2][3] The QMSR does three things:
- It incorporates ISO 13485:2016 by reference, so the standard's requirements are now FDA requirements, along with the vocabulary in clause 3 of ISO 9000:2015.[4]
- It keeps a few FDA-specific requirements where ISO 13485 doesn't go far enough for FDA, such as what complaint and servicing records must contain, recording the UDI, and labeling and packaging controls.[4]
- It drops terms ISO 13485 doesn't use. FDA said it was not keeping separate requirements for the QSR, DMR, DHF and DHR because ISO 13485 largely requires the same elements to be documented.[2]
The old-to-new translator
The four record names get the attention, but plenty of other QS regulation terms changed too. Type a term you still hear around the office.
| Old QS regulation term | What to say now | Where it lives |
|---|---|---|
| Design history file (DHF) | Design and development file | ISO 13485 7.3.10 |
| Device master record (DMR) | Medical device file | ISO 13485 4.2.3 |
| Device history record (DHR) | Record for each medical device or batch (often called the device or batch record) | ISO 13485 7.5.1; UDI recording in § 820.35(c) |
| Quality system record (QSR) | No single named record; your quality manual, procedures and records | ISO 13485 4.2.1, 4.2.2, 4.2.5 |
| Quality system | Quality management system (QMS) | ISO 13485 clause 4 |
| Management with executive responsibility | Top management | ISO 13485 clause 5; ISO 9000 definitions |
| Quality audit | Internal audit | ISO 13485 8.2.4 |
| Establish (define, document and implement) | Document. FDA says the two terms aren't identical and it didn't intend a straight swap, but under ISO 13485 a "documented" requirement must also be established, implemented and maintained.[2] | ISO 13485 0.2 |
| Corrective and preventive action (CAPA) | Corrective action and preventive action, as two separate requirements. "CAPA" is still everyday shorthand. | ISO 13485 8.5.2, 8.5.3 |
| Complaint files | Complaint handling records | ISO 13485 8.2.2; record content in § 820.35(a) |
| Acceptance activities (receiving, in-process, finished device) | Monitoring and measurement of product; verification of purchased product | ISO 13485 8.2.6, 7.4.3 |
| Production and process controls | Control of production and service provision | ISO 13485 7.5.1 |
| Inspection, measuring and test equipment | Control of monitoring and measuring equipment | ISO 13485 7.6 |
| Handling, storage, distribution | Preservation of product | ISO 13485 7.5.11 |
| Design controls | Design and development | ISO 13485 7.3; scope in § 820.10(c) |
| Device labeling, device packaging | Device labeling and packaging controls (kept as an FDA requirement) | § 820.45 |
| Servicing | Servicing activities | ISO 13485 7.5.4; record content in § 820.35(b) |
| QSIT (Quality System Inspection Technique) | Withdrawn; inspections follow Compliance Program 7382.850 | FDA inspection program |
No match in this table. Check the full crosswalk further down.
DHF → design and development file
The closest one-to-one match of the four.
Then: QS regulation
A design history file for each type of device, containing or referencing the records needed to show the design was developed according to the approved design plan and the design control requirements.[1]
Now: ISO 13485 7.3.10
One file per device type or family. It holds, or points to, the evidence that design and development requirements were met, and it must also include the records of design changes.[5]
What's the same: the file still holds or points to your design plan, inputs, outputs, reviews, verification, validation, transfer and risk management records. FDA drew the parallel itself: "consistent with the former DHF, Clause 7.3.10 requires the design and development file to contain or reference all the records necessary" to show conformity.[2]
What's different:
- Families are explicit. One file can cover a device family, not just a single device type.
- Design change records are named. The file must include records of design and development changes, which links it directly to clause 7.3.9.
- Scope is set by § 820.10(c). Design and development requirements apply to Class II and III devices and to certain Class I devices: those automated with software and a short list in the regulation.[4]
Read more: practical checks
If your files are organized by individual product, decide whether family-level files would be simpler, and write down how a family is defined.
Check that design change records are actually in, or referenced from, the file, not only in a separate change system with no link back.
Confirm the file references your risk management file, usability engineering file and software documentation, which often live elsewhere.
DMR → medical device file
Similar purpose, broader content. This is where most gaps show up.
Then: QS regulation
A device master record for each type of device: device specifications, production process specifications, quality assurance procedures and specifications, packaging and labeling specifications, and installation, maintenance and servicing procedures.[1]
Now: ISO 13485 4.2.3
One or more files per device type or family that hold, or point to, the documents showing the device meets ISO 13485 and regulatory requirements. Expected contents: what the device is, its intended use and its labeling and instructions; product specifications; how it's made, packaged, stored, handled and distributed; how it's measured and monitored; and installation and servicing requirements where relevant.[5]
What's the same: the "recipe" for building, testing, packaging and servicing the device. FDA said prior DMR content will now be found in the medical device file, and that the final design output forms the basis or starting point for it.[2]
What's different:
- It describes the device, not just how to build it. A general description, the intended use and the labeling are now expected in, or referenced from, the file.
- It reaches into storage, handling and distribution, which the DMR covered only partly.
- It has to show regulatory compliance, not only manufacturing specifications.
Read more: the EU connection
If you sell in the EU, the medical device file and the MDR technical documentation (Annexes II and III) overlap a lot, but they aren't the same thing. The technical documentation also needs, for example, the demonstration of conformity with the general safety and performance requirements (often kept as a GSPR checklist), the benefit-risk analysis, the clinical evaluation and the post-market surveillance documentation.[6]
Many companies keep the medical device file as an index that references the same underlying documents the technical documentation uses, so each document lives in one place.
DHR → record for each device or batch
Same job, shorter list in the standard. FDA's extra requirements fill the gaps.
Then: QS regulation
A device history record for each batch, lot or unit: dates of manufacture, quantity manufactured, quantity released for distribution, acceptance records, the primary identification label and labeling used, and any UDI and control numbers.[1]
Now: ISO 13485 7.5.1 plus FDA additions
One record per device or batch. It must give the traceability that 7.5.9 calls for, show how many were made and how many were approved for distribution, and be verified and approved.[5] The QMSR adds that the UDI must be recorded for each device or batch, and that labeling release is documented.[4]
What's the same: one record per unit or batch that tells the story of how it was built. FDA said many of the DHR's requirements are "largely required" to be in the device or batch record.[2] ISO 7.5.1 doesn't name dates of manufacture or the label used for each unit, but § 820.45(c) requires labeling inspection results to be documented as specified in the medical device file.[4]
Where the other DHR elements now come from:
- Acceptance records: ISO 13485 requires evidence that product meets its acceptance criteria, and the identity of the person who authorized release.[5]
- Labeling: § 820.45 requires labeling and packaging to be examined before release and the release of labeling to be documented.[4]
- UDI: § 820.35(c) requires the UDI to be recorded for each device or batch.[4]
Read more: electronic records
Many manufacturers now keep these as electronic batch or device records. The record still has to be verified and approved, so build the review and approval step into the electronic workflow rather than adding a paper signature at the end.[5]
Software that creates or manages these records is quality system software. ISO 13485 requires its application to be validated before first use and after changes, proportionate to the risk of using it.[5] FDA's electronic records and signatures rule, 21 CFR Part 11, still applies as well.[7]
QSR → your QMS documentation
The quiet one. It had no single replacement because it never had much content of its own.
Then: QS regulation
A quality system record holding, or pointing to, the procedures and records required by Part 820 that weren't specific to one device type, including management responsibility records.[1]
Now: ISO 13485 4.2
Documented quality policy and objectives, a quality manual, the procedures and records the standard requires, and the documents you need for your processes, all under document and record control.[5]
In practice, most companies' quality system record was their quality manual and procedure index. The quality manual under ISO 13485 4.2.2 does the same job, and it also has to outline the structure of your documentation.[5]
Other changes that affect your records
The record names are the visible part. These changes matter just as much on inspection day.
Management review and audit reports are now inspectable
Under the old § 820.180(c), management review, quality audit and supplier audit reports were exempt from FDA inspection; FDA could only ask for written certification that they'd been done. The QMSR removed that exception, and FDA says investigators may review records created before 2 February 2026 too.[1][2][3]
Read more
FDA's reasoning was that other regulators and MDSAP auditors already see these records, and that its investigators already had access to the underlying data.[2]
Write management review minutes and audit reports assuming an investigator will read them: factual, specific, and showing that issues raised were followed through. Records you consider confidential can be marked to help FDA decide what is protected from disclosure.[4]
QSIT is gone
FDA withdrew the Quality System Inspection Technique on 2 February 2026. Inspections now follow the updated compliance program, Inspection of Medical Device Manufacturers (7382.850).[3]
Read more
QSIT was built around the old regulation's subsystems. Compliance Program 7382.850 groups QMSR requirements into six QMS areas plus four other applicable FDA requirements, treats the quality system as a set of linked processes, and uses your risk management documentation to focus the inspection.[8]
An ISO 13485 certificate doesn't replace an FDA inspection.[2]
Complaint and servicing records have set contents
The QMSR lists what complaint and servicing records must include, such as the device name, UDI, dates and what was done.[4]
Read more
Records of review, evaluation and investigation are required for complaints involving a possible failure of the device, labeling or packaging to meet its specifications. If you don't investigate because a similar complaint was already investigated, record why. For complaints reportable under FDA's medical device reporting rule, and for any complaint you investigate, § 820.35(a) lists seven items the record must contain.[4]
Check your complaint and service templates field by field against § 820.35(a) and (b). Missing fields are an easy finding to avoid.
Labeling and packaging controls stayed
FDA kept its own labeling and packaging requirements: examining labeling for accuracy before release, documenting release, and preventing mix-ups.[4]
Read more
The examination covers items such as the UDI, expiration date, storage and handling instructions, and any processing instructions.[4]
If you followed the old labeling sections closely, little changes here beyond the section number.
The vocabulary follows ISO
ISO 13485 and ISO 9000 definitions apply, with a handful of FDA definitions kept, such as component, finished device, manufacturer, remanufacturer and rework.[4]
Read more
Risk management runs through everything
ISO 13485 expects a risk-based approach across the quality system, not only in design.[5]
Read more
The old regulation mentioned risk analysis only once in its text, in design validation (§ 820.30(g)), although FDA expected it throughout design. ISO 13485 applies a risk-based approach to controlling QMS processes and requires documented risk management in product realization.[5]
Expect questions about how risk shaped decisions in purchasing, process validation, nonconformance and complaint handling.
Do you have to rename your files?
Nothing in the final rule requires you to rename records. FDA noted that each organization will implement a quality system specific to its needs, and what it requires is that records are kept to meet the QMSR.[2] There are good reasons to update the names anyway, and a sensible middle path.
| Keep old names, add a crosswalk | Rename to ISO terms | |
|---|---|---|
| Effort now | Low: one mapping table in the quality manual | Higher: procedures, templates, software and training |
| Risk of confusion | Ongoing, especially with new hires and auditors used to ISO terms | Short-term, during the transition |
| Gap check | Easy to skip, because nothing looks different | Forced, because each record is re-examined |
| Best for | Stable products with few changes planned | Companies already updating procedures, systems or selling outside the US |
Whichever you choose, do the gap check. The content of the medical device file and the device or batch record is where real differences hide, and renaming a DMR "medical device file" without adding the device description, intended use and labeling doesn't make it one.
A practical transition checklist
If you haven't finished the move, or want to check it, work through these.
- Write the crosswalk. Map every old record name and procedure to its QMSR and ISO 13485 home, and put it in the quality manual.
- Gap-check each medical device file. Device description, intended use, labeling and instructions, storage, handling and distribution, and installation and servicing, as applicable.
- Gap-check each design and development file. Family scope defined, design change records included or referenced, risk and usability files referenced.
- Gap-check device and batch records. Traceability, quantity made and approved for distribution, verification and approval, UDI, acceptance and release.
- Update complaint and servicing templates against the field lists in § 820.35(a) and (b).
- Review how you write management review and audit reports, now that FDA can read them.
- Update procedures, forms and software labels so the vocabulary is consistent, or so the crosswalk explains every difference.
- Train people, especially on what changed in content, not just in names.
- Run an internal audit against the QMSR and the clauses it points to, using ISO 13485 terms.
The full crosswalk: old sections to their new home
For reference, here is where the main sections of the former QS regulation now live. It's our working map, not an official FDA table, and several old sections spread across more than one ISO clause. Always check the full text.[1][5][4]
Show the crosswalk (32 rows)
| Former QS regulation | Now |
|---|---|
| 820.1 Scope | § 820.1 |
| 820.3 Definitions | ISO 13485 clause 3, ISO 9000 clause 3 and § 820.3 |
| 820.5 Quality system | 4.1 General requirements; § 820.10(a) |
| 820.20 Management responsibility | ISO 13485 clause 5 |
| 820.22 Quality audit | 8.2.4 Internal audit |
| 820.25 Personnel | 6.2 Human resources |
| 820.30 Design controls | 7.3 Design and development; scope in § 820.10(c) |
| 820.30(j) Design history file | 7.3.10 Design and development files |
| 820.40 Document controls | 4.2.4 Control of documents |
| 820.50 Purchasing controls | 7.4 Purchasing |
| 820.60 Identification | 7.5.8 Identification; UDI in § 820.10(b) |
| 820.65 Traceability | 7.5.9 Traceability; § 820.10(b) and (d) |
| 820.70 Production and process controls | 7.5.1, 7.5.2 Cleanliness of product, 6.3 Infrastructure, 6.4 Work environment and contamination control; automated processes (820.70(i)) in 4.1.6 and 7.5.6 |
| 820.30(g) Risk analysis in design validation | 7.1 Planning of product realization (risk management); 7.3.7 |
| 820.72 Inspection, measuring and test equipment | 7.6 Control of monitoring and measuring equipment |
| 820.75 Process validation | 7.5.6 (and 7.5.7 for sterilization and sterile barrier processes) |
| 820.80 Acceptance activities | 8.2.6 Monitoring and measurement of product; 7.4.3 Verification of purchased product |
| 820.86 Acceptance status | 7.5.8 Identification (status) |
| 820.90 Nonconforming product | 8.3 Control of nonconforming product |
| 820.100 Corrective and preventive action | 8.5.2 Corrective action, 8.5.3 Preventive action |
| 820.120 Device labeling | § 820.45 |
| 820.130 Device packaging | § 820.45; 7.5.11 Preservation of product |
| 820.140, 820.150 Handling, storage | 7.5.11 Preservation of product |
| 820.160 Distribution | 7.5.11; 7.5.9 Traceability |
| 820.170 Installation | 7.5.3 Installation activities |
| 820.180 Records, general | 4.2.5 Control of records; § 820.35 |
| 820.181 Device master record | 4.2.3 Medical device file |
| 820.184 Device history record | 7.5.1 record for each device or batch; § 820.35(c); § 820.45 |
| 820.186 Quality system record | 4.2.1, 4.2.2 Quality manual, 4.2.5 |
| 820.198 Complaint files | 8.2.2 Complaint handling; § 820.35(a); § 820.10(b) |
| 820.200 Servicing | 7.5.4 Servicing activities; § 820.35(b) |
| 820.250 Statistical techniques | 8.1, 8.4 Analysis of data; sample size rationale in 7.3.6, 7.3.7 |
Common questions
Is it wrong to still say "DHF"?
Not wrong in conversation, and plenty of people will for years. In procedures, submissions and anything an investigator reads, use the current terms or explain the mapping.
Read more
You'll still see the old terms in older FDA guidance documents written under the QS regulation, and in some software tools. That doesn't make them current regulatory terms.
The risk with old names isn't the word itself. It's assuming that because the "DHF" exists, the design and development file requirements are met, without checking what changed.
Is the medical device file the same as the EU technical documentation?
No. They overlap heavily, but the EU technical documentation asks for more, such as the clinical evaluation and the demonstration of conformity with the general safety and performance requirements (often kept as a GSPR checklist).[6]
Read more
A common approach is to make the medical device file an index that references the same controlled documents used in the technical documentation, so nothing is duplicated.
Keep both current from the same change process, so a change updates every file that references the changed document.
Do we need a separate file for every device?
No. ISO 13485 allows the design and development file and the medical device file to cover a device type or a device family.[5]
Read more
ISO 13485 defines a family as devices sharing the same basic design and performance characteristics related to safety, intended use and function.[5] Write down how you apply that, for example variants that differ only in size or configuration.
Family-level files reduce duplication, but each variant's specific documents still need to be identifiable within the file.
Does our ISO 13485 certificate mean we comply with the QMSR?
Not automatically. The QMSR adds FDA-specific requirements, and FDA has said ISO 13485 certificates won't replace its inspections.[2]
Read more
The FDA additions include complaint and servicing record contents, UDI, labeling and packaging controls, and links to other FDA rules such as medical device reporting and corrections and removals.[4]
A certified ISO 13485 system is a strong starting point. Check the FDA-specific sections and the inspection program on top of it.
Do Class I devices need a design and development file?
Only if design and development requirements apply to them: Class I devices automated with software and those on the regulation's short list.[4]
What happened to records from before February 2026?
They don't need to be rewritten. Records created under the QS regulation stay valid records of what was done at the time, and FDA investigators may review them.[3]
Read more
What changes is how you maintain the files going forward. When you next update a product's files, bring them into the new structure and fill any gaps in content.
Keep them retrievable for the lifetime of the device as you define it, and at least two years from release, as ISO 13485 requires.[5]
References
- U.S. FDA, Quality System Regulation, 21 CFR Part 820 (as in force until 1 February 2026), including §§ 820.3, 820.30(j), 820.181, 820.184 and 820.186; final rule published at 61 FR 52602, 7 October 1996, as amended, including the UDI rule, 78 FR 58786, 24 September 2013. www.govinfo.gov/content/pkg/FR-1996-10-07/html/96-25720.htm
- U.S. FDA, Medical Devices; Quality System Regulation Amendments, final rule, 89 FR 7496, 2 February 2024 (preamble, including the response to Comment 31 on the QSR, DMR, DHF and DHR, and the removal of the former § 820.180(c) exception). www.federalregister.gov/documents/2024/02/02/2024-01709/medical-devices-quality-system-regulation-amendments
- U.S. FDA, Quality Management System Regulation: Frequently Asked Questions, including the effective date, inspection of management review and audit records, and the withdrawal of QSIT in favour of Compliance Program 7382.850. www.fda.gov/medical-devices/quality-management-system-regulation-qmsr/quality-management-system-regulation-frequently-asked-questions
- 21 CFR Part 820, Quality Management System Regulation, §§ 820.1, 820.3, 820.7, 820.10, 820.35 and 820.45, effective 2 February 2026. www.ecfr.gov/current/title-21/chapter-I/subchapter-H/part-820
- ISO 13485:2016, Medical devices — Quality management systems — Requirements for regulatory purposes, clauses 0.2, 3.12, 4.1, 4.1.6, 4.2.1 to 4.2.5, 7.1, 7.3.6, 7.3.7, 7.3.9, 7.3.10, 7.5.1, 7.5.6, 7.5.7, 7.5.9, 8.2.2, 8.2.4, 8.2.6, 8.5.2 and 8.5.3, and other clauses as cited. www.iso.org/standard/59752.html
- Regulation (EU) 2017/745 (EU MDR), Annexes II and III, technical documentation and technical documentation on post-market surveillance. eur-lex.europa.eu/eli/reg/2017/745/oj
- 21 CFR Part 11, Electronic Records; Electronic Signatures. www.ecfr.gov/current/title-21/chapter-I/subchapter-A/part-11
- U.S. FDA, Compliance Program 7382.850, Inspection of Medical Device Manufacturers, effective 2 February 2026 (supersedes CP 7382.845). www.fda.gov/media/80195/download
- U.S. FDA, Quality Management System Regulation (QMSR) web page, including the note that exemption from CGMP requirements does not exempt manufacturers from complaint files or general records requirements under 21 CFR 820.35. www.fda.gov/medical-devices/postmarket-requirements-devices/quality-system-qs-regulationmedical-device-good-manufacturing-practices
Regulatory text was checked against the Federal Register, eCFR and FDA. The crosswalk is a working map prepared for this article, not an official FDA table; check the full text of the regulation and ISO 13485.